Philippine Journal of Pathology 2025;10(2):26-32

doi:https://doi.org/10.21141/PJP.2025.08

Quantifying total allowable error violations in serum-sodium quality control: A computer simulation experiment of two- to six-sigma processes.

Mark Angelo ANG 1 ; Karen Cybelle SOTALBO 1

Affiliations

+expand

Keywords

Rstudio; Total Allowable Error

Country

Philippines

Language

Abstract

BACKGROUND

Serum-sodium reporting tolerates a total allowable error (TEa) of only ±4 mmol/L, yet many laboratories continue to operate at the marginal three-sigma level because the quantitative benefit of additional sigma capability is poorly characterized.

OBJECTIVES

The study aims to translate sigma metrics into clinically intuitive risk estimates by (1) quantifying the proportion of QC results that exceed the TEa at five sigma levels (2 – 6 σ) and (2) determining whether successive sigma gains produce statistically significant reductions in error.

METHODOLOGY

Five (5) hypothetical assays were parameterized with a common mean of 140 mmol/L and CVs corresponding to 2-, 3-, 4-, 5- and 6-sigma performance. For each assay, 1,000 Monte-Carlo iterations were run, each iteration simulating 36,500 QC results (assuming 100 runs/day for 365 days) drawn from N(μ = 140, σ = μ × CV). The error rate (the proportion of results outside ±4 mmol/L) was recorded per iteration. Distributions were summarized (mean, range, SD); differences were evaluated with one-way ANOVA followed by Tukey’s HSD.

RESULTS

Mean (±SD) error rates declined significantly with increasing sigma: Assay A (2 σ): 0.0456 ± 0.0011; Assay B (3 σ): 0.00270 ± 0.00027;Assay C (4 σ): 6.3 × 10-5 ± 4.1 × 10-5; Assay D (5 σ): 5.8 × 10-7 ± 8.0 × 10-7; and Assay E (6 σ): 2.0 × 10-7 ± 3.1 × 10-7. The maximum single-iteration error rate fell from 0.0505 at 2 σ to 1.1 × 10-4 at 4 σ. The 5 σ and 6 σ processes produced zero TEa violations in ≥96 % of iterations. ANOVA confirmed a global difference (p < 0.001); all pairwise contrasts were significant (p < 0.001) except between 5 σ vs 6 σ (p = 0.62).

NEW ABSTRACT STRUCTURE 1

Each one-sigma gain yields an order-of-magnitude reduction in TEa violations until a plateau is reached at ≥5 σ, where residual analytical risk is negligible. These simulations support the recommendation that laboratories operating serum-sodium assays below 4 σ should prioritize precision improvements or enhanced QC strategies, whereas ≥5 σ assays may safely adopt less intensive QC without compromising patient safety.