Journal of Pharmaceutical Practice and Service 2025;43(12):599-602

doi:10.12206/j.issn.2097-2024.202311012

Effect of different conditions of solid phase extraction column on determination of morphine in compound liquorice oral solution

Linlin LIU 1 ; Guixiang ZHONG 1 ; Jia YAN 1 ; Xin ZHOU 1

Affiliations

+expand

Keywords

solid phase extraction column; compound liquorice oral solution; morphine

Country

China

Language

Chinese

Abstract

Objective To explore the effect of different packing contents of solid phase extraction column, repeated use times and solid phase extraction column under pressure on the determination of morphine content in compound liquorice oral solution. Methods The component was performed by high performance liquid chromatography (HPLC) equipped with the C8 (4.6 mm × 150 mm, 5 μm) chromatographic column. The mobile phase consisted of acetonitrile-2.5 mmol/L sodium heptanesulfonate solution-50 mmol/L potassium dihydrogen phosphate solution (5∶18∶18). The flow rate was 1.0 ml/min with the 220 nm detection wavelength, 20 μl injection volume and 30℃ column temperature. Results It was found that the SPE column produced of 500 mg had a higher recovery rate of morphine, and the RSD value was small. It was found that SPE column could be reused for up to 3 times under the condition of guaranteeing recovery rate.In the pressurized experiment,at the flow rate of 0.5 ml/min, the measured morphine content was basically consistent with that measured at atmospheric pressure. Conclusion The determination of morphine in compound liquorice oral solution could be carried out by C18 solid phase extraction column of 500 mg. Under pressure, the flow rate was about 0.5 ml/min for morphine recovery, which could achieve high recovery of morphine in compound liquorice oral solution and improve the detection efficiency of morphine.