Chinese Journal of Radiological Health 2025;34(3):348-353
doi:10.13491/j.issn.1004-714X.2025.03.007
Analysis of differences in radiation dose rates in the fluoroscopic protection zone of digital subtraction angiography devices
Hongwei YU 1 ; Zhan TAN 1 ; Pengxiang QU 1 ; Weixu HUANG 1 ; Xuan LONG 1
Affiliations
Keywords
Radiation protection; Interventional radiology; Radiation dose rate; Operator
Country
China
Language
Chinese
Abstract
Objective To analyze bedside radiation dose rates for interventional surgery operators in Guangdong Province, examine dose distribution patterns, and identify potential weaknesses in radiation protection, and to provide guidance for optimizing radiation monitoring and protection measures. Methods A total of 209 digital subtraction angiography devices measured in Guangdong Province between 2017 and 2024 were used as the research objects. The first and second operator positions were set at 30 cm and 90 cm horizontally from the X-ray tube focal point, respectively. Monitoring points were set up at 155, 125, 105, 80, and 20 cm above the ground. Results The median bedside radiation dose rate for interventional surgery operators in Guangdong Province was 83.0 (3.9,