Chinese Journal of Radiological Health 2024;33(6):675-680

doi:10.13491/j.issn.1004-714X.2024.06.010

Analysis of issues in quality control test results of linear accelerators in China

Hongbo WANG 1 ; Xuexia LIU 2 ; Songhui ZHANG 1 ; Chenxu HUANG 1 ; Yuexin GUO 1 ; Quanfu SUN 3

Affiliations

+expand

Keywords

Linear accelerator; Quality control test; Unqualified item; Investigation of current situation

Country

China

Language

Chinese

Abstract

Objective To study the unqualified items in the reported quality control tests of linear accelerators, analyze the issues in quality control tests, and propose the key points and development directions for accelerator quality control test in China. Methods A literature review was conducted using the CNKI database to analyze the qualified rates of test items and the issues in quality control tests. Results In the literature on the quality control tests of linear accelerators, except for a few provinces where the qualified rates of all test items were 100%, unqualified items were reported in most of the literature. There were unqualified items related to X-ray and electron beam in different reports. Error of dose indication was the unqualified item with the highest occurrence rate in X-ray test, and the item with the lowest qualified rates in X-ray and electron beam tests. The lowest qualified rate of X-ray dose indication error was 73.5% in 2016, and the lowest qualified rate of electron beam dose indication error was 46.2% in 2017. Conclusion Tests should be carried out strictly according to the items and intervals specified by the quality control test standards. Hospitals, radiation health technology service institutions, and health administrative departments should each fulfill their respective responsibilities, work together, and place emphasis on ensuring effective quality control tests of linear accelerators to further enhance the overall quality control standards for these devices.