Risk Control of UPS Power Supply for Medical Equipment.
10.3969/j.issn.1671-7104.2021.04.024
- Author:
Hao YAN
1
;
Heqing LU
1
;
Qi SU
1
Author Information
1. Medical Equipment Department, Shanghai First Maternity and Infant Hospital, School of Medicine, Tongji University, Shanghai,
- Publication Type:Journal Article
- Keywords:
Internet of Things;
risk control;
uninterruptible power supply
- MeSH:
Electric Power Supplies;
Hospitals;
Humans;
Technology
- From:
Chinese Journal of Medical Instrumentation
2021;45(4):459-462
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:Putting forward that the risk of using UPS for medical matching is easy to be ignored by managers.
METHODS:The method of UPS inspection was designed and used in hospital.
RESULTS:The qualified rate of UPS for equipment matching was 66.7% and the problems were corrected. Based on the Internet of Things technology, the UPS monitoring technology was prospected.
CONCLUSIONS:It is effective to control the risk of UPS by recognizing the risk of using UPS and specifying the strategy.