Validity study of Coda Motion Analysis System for measuring cervical lateral flexion in normal adults.
10.3969/j.issn.1003-0034.2018.08.007
- Author:
Hui SONG
1
;
Zhong-Yang GAO
1
;
Ying-Jie ZHAO
1
;
Zheng-Chao GAO
1
;
Xiao-Wen QIU
1
;
Hao-Peng LI
1
;
Xi-Jing HE
2
Author Information
1. Department of Orthopaedics, the 2nd Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710004, Shaanxi, China.
2. Department of Orthopaedics, the 2nd Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710004, Shaanxi, China; xijing_h@vip.tom.com.
- Publication Type:Journal Article
- Keywords:
Cervical range of motion;
Coda Motion Analysis System;
Cortical bone trajectory;
Lateral flexion
- MeSH:
Adult;
Cervical Vertebrae;
Eye Diseases, Hereditary;
Humans;
Neck;
Range of Motion, Articular;
Reproducibility of Results
- From:
China Journal of Orthopaedics and Traumatology
2018;31(11):1017-1021
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To investigate the validity of Coda Motion Analysis System for measuring cervical lateral flexion in normal adults in order to explore a new measuring tool for clinical and research practice.
METHODS:A total of 43 participants were involved in the study. Cervical range of lateral flexion were measured with Coda Motion Analysis System and "gold standard" X-ray simultaneously. The validity and agreement were assessed using the scatter diagram, the Pearson correlation coefficient and limits of agreement.
RESULTS:Cervical range of lateral flexion measured by Coda Motion Analysis System had no statistical differences with those measured by X-ray(>0.05). The Coda Motion Analysis System demonstrated a very good linear relation with the X-ray measurements in cervical range of right lateral flexion, left flexion and total lateral flexion, and the Pearson correlation coefficients were 0.72, 0.85 and 0.90 respectively.
CONCLUSIONS:Coda Motion Analysis System showed good validity for measuring cervical lateral flexion in normal adults. Because the reliability of Coda Motion Analysis System was established previously, the results of this study suggest that the system has the potential to be used to measure cervical lateral flexion in clinical and research practice.