Assessment of uncertainty in Hp(10) measurement by using a double TLD element personal dosimeter
10.3760/cma.j.issn.0254-5098.2019.05.014
- VernacularTitle:双热释光元件个人剂量计Hp(10)测量不确定度评定
- Author:
Xuan ZHANG
1
;
Wen GUO
;
Yanqiu DING
;
Aiying HU
;
Fei CHEN
Author Information
1. 中国疾病预防控制中心辐射防护与核安全医学所辐射防护与核应急中国疾病预防控制中心重点实验室
- Keywords:
Hp(10);
Uncertainty of measurement;
GUM method
- From:
Chinese Journal of Radiological Medicine and Protection
2019;39(5):397-400
- CountryChina
- Language:Chinese
-
Abstract:
Objective To describe the evaluation method of uncertainty in measurement of depth individual dose equivalent.Methods Based on the performance of the personal dosimeter,GUM method was used to evaluate the uncertainty of Hp(10) measurement by using a personal dosimeter.Results The relative extended uncertainty was 28% when the measurement result was corrected by nonlinear response and energy response,as well as 33% when not corrected in the same way.Conclusions This method can be used to effectively evaluate the uncertainty arising from personal dose equivalent Hp(10) measurement in routine monitoring.