The design of circuit for detecting multi-physiological signals.
- Author:
Jie LIU
1
;
Bo WANG
;
Wen LIU
Author Information
1. School of Computer and Information Technology, Beijing Jiaotong University, Beijing 100044, China. jieliu@bjtu.edu.cn
- Publication Type:Journal Article
- MeSH:
Arteriosclerosis;
diagnosis;
Blood Pressure;
physiology;
Equipment Design;
methods;
Heart Sounds;
physiology;
Humans;
Monitoring, Physiologic;
instrumentation;
methods;
Pulse;
Signal Processing, Computer-Assisted
- From:
Journal of Biomedical Engineering
2009;26(6):1230-1233
- CountryChina
- Language:Chinese
-
Abstract:
In this paper is presented the design of detection circuit and single chip acquisition circuit for heart sound, blood pressure and pulse wave. The multi-physiological signals from single chip are synthesized and processed by master computer. The master computer acquires, analyzes and displays the multi-signals by Software LabVIEW. The three physiological signals are fused to calculate the artery stiffness index by time relation of multi-physiological signals. The experiment shows that the circuit is reliable and effective so that it can be used for human vessel function evaluation.