Application of artificial intelligence in quality control of mammographic images
10.13491/j.issn.1004-714X.2026.02.003
- VernacularTitle:人工智能在乳腺X射线摄影图像质量控制中的应用
- Author:
Yunyun LYU
1
;
Le FU
1
;
Ruixin LI
1
;
Zeyi ZHANG
1
;
Xiaoli MU
1
;
Hui WANG
1
;
Huizhi CAO
2
;
Jianli YU
1
Author Information
1. Department of Radiology, Obstetrics and Gynecology Hospital of Tongji University, Shanghai 201204, China.
2. General Electric Medical X-ray Imaging Collaboration Center, Beijing 100176, China.
- Publication Type:OriginalArticles
- Keywords:
Artificial intelligence;
Mammography;
Image quality control;
Consistency assessment
- From:
Chinese Journal of Radiological Health
2026;35(2):173-179
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the application value of artificial intelligence (AI) in the quality control of mammographic images and explore its feasibility for improving image quality. Methods A retrospective analysis was conducted on 500 mammographic images from 125 female patients. These images were acquired in December 2024 at the Department of Radiology, Obstetrics and Gynecology Hospital of Tongji University by two junior technologists with qualification certificate for junior radiologic technologists and ≤ 2 months of independent operation. The reference standard was the evaluation by a panel of senior experts comprising one associate chief radiologist and two intermediate radiologic technologists with over 10 years of experience in mammography. The evaluation was based on nine criteria within an AI quality control system. The correlation and consistency (Kappa test) of the assessment outcomes were compared among the junior technologist group, the AI quality control group, and the panel group. Additionally, the differences in the proportions of high-, medium-, and low-quality images rated by the three groups were analyzed. After one month of AI assistance, 200 images from 50 additional patients acquired by the same two junior technologists in January 2025 were collected. The area under the receiver operating characteristic curve and 95% confidence interval (95%CI) were calculated for the assessment performance of the junior technologist group before and after AI assistance. Differences were compared using the DeLong test. Changes in the disqualification rates for the nine criteria were analyzed. Results In consistency analysis, the AI quality control group showed high consistency with the panel group across multiple key criteria, with Kappa values ranging from 0.41 to 1.00. In contrast, the consistency between the junior technologist group and the panel group was generally low, with Kappa values ranging from 0.13 to 0.49. In comparison of image quality classification, no significant differences were observed in image quality classification between the AI group and the panel group (P>0.05). However, the proportion of images rated as high quality by the junior technologist group was significantly higher, while the proportions rated as medium and low quality were significantly lower, compared to those rated by the panel group (P<0.05). After AI assistance, the area under the receiver operating characteristic curve for the assessment performance of junior technologist group increased significantly from 0.56 (95%CI: 0.51-0.61) to 0.91 (95%CI: 0.87-0.94) (P<0.001). The disqualification rates for skin folds, incomplete inclusion of the pectoralis major muscle, and nipple not in profile decreased significantly. Conclusion AI demonstrates assessment capability comparable to the panel group in mammographic quality control. AI effectively enhances the quality of images acquired by junior technologists through real-time, objective feedback.