Retrospective evaluation of mandibular third molars using simulated low-dose cone-beam computed tomography: A comparative image quality study
- Author:
Dalia Homar ASAN
1
;
Josefine CEDERHAG
;
Xie-Qi SHI
;
Mats NILSSON
;
Kristina HELLÉN-HALME
Author Information
- Publication Type:Original Articles
- From:Imaging Science in Dentistry 2026;56(1):54-61
- CountryRepublic of Korea
- Language:English
-
Abstract:
Purpose:This study was performed to establish a procedure for simulated low-dose cone-beam computed tomography (CBCT) scans and to investigate whether the resulting images are comparable in diagnostic accuracy to those obtained using a clinical low-dose protocol.
Materials and Methods:ImageJ was used to manipulate the sinogram data from CBCT scans acquired at 5 mA to mimic a technical setting of 2 mA by adding noise to the Radon-transformed projection data before image reconstruction. Four observers compared the simulated 2 mA CBCT scans with original clinical 2 mA CBCT scans acquired previously. The CBCT images were analysed using a protocol with a ranking scale, and the observers were required to select only 1 category for each variable. The Wilcoxon signed-rank test was used to assess differences between the 2 CBCT scan types, with a significance level of P<0.05. Intra-observer agreement was evaluated using the Cohen kappa.
Results:Pairwise observer comparisons of the simulated and clinical low-dose CBCT scans showed no significantdifferences in image quality. Intra-observer agreement was acceptable, and in 5 comparisons, the results indicated a high degree of agreement.
Conclusion:Simulated low-dose CBCT scans can be generated using ImageJ. No significant differences in imagequality were observed between simulated and clinical low-dose CBCT scans when evaluating mandibular third molars. These findings suggest that manipulation of sinogram data is a promising radiation-free approach for simulating low-dose images in optimisation efforts.