Phase Ⅰ clinical study of bilateral catheter-based ultrasound renal denervation in patients with uncontrolled hypertension
10.3760/cma.j.cn112148-20250715-00510
- VernacularTitle:双面导管超声RDN系统治疗药物控制不佳高血压患者Ⅰ期临床研究
- Author:
Jun QIAN
1
;
Yun DU
;
Wei SHA
;
Shunkang RONG
;
Yuanqing YAO
;
Rong HU
;
Changming DENG
;
Dichuan LIU
;
Jianlin DU
;
Qiang SHE
;
Bo YU
;
Xiaoping CHEN
;
Jing HUANG
Author Information
1. 重庆医科大学附属第二医院心血管内科,重庆 400010
- Publication Type:Journal Article
- Keywords:
Hypertension;
Renal denervation;
Ultrasound;
Four-quadrant;
Ablation
- From:
Chinese Journal of Cardiology
2025;53(11):1225-1230
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To preliminarily evaluate the efficacy and safety of a domestically developed bilateral interventional ultrasound renal denervation (RDN) system in patients with uncontrolled hypertension despite antihypertensive medication.Methods:A multicenter, single-arm trial was conducted. Patients with uncontrolled hypertension (≥2 antihypertensive drugs) were enrolled from April 2023 to April 2024 at the Second Affiliated Hospital of Chongqing Medical University, West China Hospital of Sichuan University, and the Second Affiliated Hospital of Harbin Medical University. RDN was performed using the UltraCure? bilateral interventional ultrasound system via femoral or brachial artery access. Multi-segmental "quadrant-based" ablation was performed in bilateral main renal arteries and branches/accessory arteries (diameter≥4 mm). Primary endpoints were changes in office systolic blood pressure (SBP) and 24-hour daytime SBP at 2-and 6-months post-procedure. The primary safety endpoints included the incidence of major adverse events, device-related adverse events, and puncture site complications.Results:Ten patients, mean aged 47.1 years, including 9 male, successfully completed RDN. At 2 and 6 months post-procedure, office SBP decreased by (19.7±15.2) mmHg ( P=0.002, 1 mmHg=0.133 kPa) and (13.8±13.9) mmHg ( P=0.013) from baseline, while the 24-hour daytime SBP decreased by (13.4±10.6) mmHg ( P=0.004) and (11.2±9.2) mmHg ( P=0.004). Apart from one case of a limited distal renal artery dissection, no other serious device/procedure-related adverse events were observed. At 6-month follow-up, the estimated glomerular filtration rate remained stable ((85.3±18.3) ml·min -1·1.73 m -2 vs. (82.3±19.2) ml·min -1·1.73 m -2, P=0.41). No renal artery stenosis was detected. Conclusions:The domestic interventional ultrasound RDN system could effectively reduce office and ambulatory blood pressure in patients with uncontrolled hypertension, demonstrating a favorable safety profile. Long-term efficacy requires confirmation through large-scale randomized controlled trials.