Value analysis of management model of data mining in reducing failure rate of medical imaging equipment
10.3969/j.issn.1672-8270.2025.05.024
- VernacularTitle:数据挖掘管理模型在降低医疗影像设备故障率中的价值分析
- Author:
Peng ZHOU
1
;
Qiong LIU
;
Wenfei XING
;
Chaozhi ZHANG
;
Yuying YAO
Author Information
1. 中国人民解放军总医院海南医院医学工程科 三亚 572013
- Publication Type:Journal Article
- Keywords:
Data mining;
Management model;
Imaging equipment;
Failure rate;
Quality control
- From:
China Medical Equipment
2025;22(5):121-126
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To construct a management model of data mining for medical imaging equipment,so as to improve the quality of managing equipment.Methods:A management model of mining data was constructed to manage medical imaging equipment.A total of twenty imaging equipment that were using at Hainan Hospital of the General Hospital of the People's Liberation Army of China from April 2022 to March 2024 were selected.According to different management methods,the conventional management was adopted to manage them during April 2022 to March 2023,and the management model of mining data(model management)was adopted to manage them during April 2023 to March 2024.A self-developed questionnaire was used to conduct a satisfaction survey for imaging physicians,staffs of operating and maintaining equipment,and technicians who using and managing equipment,and patients who received diagnosis and treatment by using equipment.The failure rate and the imaging effect of equipment,the satisfaction scores of the relative staffs who used equipment,and the growth amplitude of operational benefits of equipment between two management methods were compared.Results:A total of 20 failures occurred in imaging equipment that adopted model management.In them,the failure rates of the self-equipment,improper operation and insufficient professional level were respectively 15%,5%and 5%,all of which were lower than those of the conventional management method.The predicted failure rate of model management was 75%,which was higher than that of the conventional management method,and the differences of the above indicators between two methods were statistically significant(x2=6.547,4.392,5.124,6.701,P<0.05).The scores of image clarity,qualification rate,excellent rate,qualification rate of body position,and the total score of imaging effect of adopting model management method were respectively(22.36±2.01),(23.21±1.54),(22.65±1.87),(23.21±1.52)and(91.43±6.77)points,all of which were higher than those of the conventional management method,and the differences were statistically significant(t=10.662,12.727,15.324,16.333,13.742,P<0.05).The satisfaction scores of imaging physicians,staffs of operation and maintenance,technicians and patients who uses management for adopting the model management method were all higher than those of the conventional management method,and the differences were statistically significant(t=13.586,14.249,17.021,11.006,P<0.05).The average values of growth amplitude of cost and benefit of equipment operation of adopting model management method were higher than those of the conventional management method,and the average value of the growth amplitude of the cost of the troubleshooting of model management method was lower than that of the conventional management method,and the differences of them were statistically significant(t=15.057,19.310,18.336,P<0.05).Conclusion:The application of the management model of data mining for medical imaging equipment can provide warning of equipment failures in advance,and reduce the failure rate of equipment,and improve the quality of management and operation of equipment,and enhance the service level of equipment.