Study on management mode of telemetry monitoring system based on FMEA and WBS
10.3969/j.issn.1672-8270.2025.01.020
- VernacularTitle:基于故障模式与影响分析和工作分解结构方法的遥测监护系统管理模式研究
- Author:
Junlin HOU
1
;
Hongyan ZHU
;
Ying XU
;
Tianbai QI
Author Information
1. 湖州市第三人民医院医学工程科 湖州 313000
- Publication Type:Journal Article
- Keywords:
Failure mode and effects analysis (FMEA);
Work breakdown structure (WBS);
Management mode;
Telemetry monitoring;
Quantitative risk score
- From:
China Medical Equipment
2025;22(1):108-113
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To explore the application value of the combination of failure mode and effects analysis (FMEA) and work breakdown structure (WBS) on management of telemetry monitoring system. Methods:Seven sets of used telemetry monitoring systems in Huzhou Third People's Hospital from March 2022 to April 2024 were selected. The routine mode was used to manage the telemetry monitoring systems from March 2022 to February 2023,and the management mode of telemetry monitoring systems based on FMEA and WBS was adopted to manage the telemetry monitoring systems from May 2023 to April 2024. FMEA method was used to analyze the quantitative risk scores of various failures in 1740 uses of the telemetry monitoring system from March 2022 to February 2023,and the fault type was represented by Fx. According to the risk quantification score,the top 5 failure types included poor radio frequency (RF) signal (F2),larger pseudo-interference (F3) of electrocardiogram (ECG),abnormal measurement of blood oxygen saturation (SpO2) (F6),excessive movement and mistake touch (F8) of patients,and identification (ID) number error (F9) of telemetry emission box. WBS quality management method combined with tree chart,countermeasure fishbone chart,task assignment matrix and other tools were performed to conduct the setting of expected objectives,countermeasure analysis,task formulation and work implementation of management of telemetry monitoring system. The above indicators were compared with the changes of the top 5 failure risk grading and incidence of failure in 1850 uses of management of telemetry monitoring system from May 2023 to April 2024. Results:After the telemetry monitoring system was used to conduct management,the risk grading of the F2,F3,F6,F8 and F9 failures with highest risk grading reduced respectively to low grade,middle grade,middle grade,low grade and middle grade. At the same time,the number of F2,F3,F6,F8,F9 failures with the highest risk grading and total failures were respectively 91,89,39,45,39 and 478 times,and the incidence rates of them were 4.92%,4.81%,2.11%,2.43%,2.11% and 25.84%,which were significantly lower than those of adopting conventional management methods. The differences of them between two methods were statistically significant (x2=2.125,6.911,9.099,2.261,2.431,32.491,P<0.05). Conclusion:The management of combined quality management method of FMEA and WBS for telemetry monitoring system can reduce risk grading and incidence rate of equipment failure,and ensure the safety of equipment in clinical use.