3D C-arm Combined With Orthopedic Robot-assisted Iliosacral Screw Fixation Through S3 Vertebra for Pelvic Fracture:Report of 4 Cases
10.3969/j.issn.1009-6604.2025.10.011
- VernacularTitle:三维C臂结合骨科手术机器人辅助经S3椎体骶髂螺钉置入治疗骨盆骨折4例
- Author:
Zongdong ZHU
1
;
Bo TAN
1
;
Zhi TANG
1
;
Jiang HU
1
;
Dan WEI
1
;
Feng LIAO
1
Author Information
1. 四川省医学科学院·四川省人民医院骨科 电子科技大学医学院,成都 610072
- Publication Type:Journal Article
- Keywords:
Pelvic fracture;
Iliosacral screw;
Third sacral vertebra;
Three-dimensional imaging;
Robot-assisted orthopedic surgery
- From:
Chinese Journal of Minimally Invasive Surgery
2025;25(10):636-640
- CountryChina
- Language:Chinese
-
Abstract:
Objective To verify the feasibility and safety of three-dimensional(3D)image acquisition by C-arm combined with orthopedic robot-assisted iliosacral screw fixation through S3 vertebra.Methods From July 2023 to June 2024,a total of 4 patients with closed pelvic fracture were surgically treated.Preoperative pelvic CT 3D reconstruction showed the presence of an ideal iliosacral screw bone channel through S3 vertebra which required screw fixation.There was no such channel through the S1 vertebra,or the stability of fixation through the S3 channel was better than that through the S1 channel.The 3D image acquisition by C-arm combined with orthopedic surgical robot was used to insert screws.Follow-ups were conducted for at least 6 months after surgery to assess fracture healing and limb function.Results All the 4 patients were successfully implanted with a S2 and a S3 iliosacral screw,and all the 8 screws did not penetrate the cortical bone.Three cases had excellent Majeed scores and 1 case had good scores at 3 months after surgery.Follow-ups were conducted for 18,12,7,and 6 months after surgery,respectively,and all fractures healed smoothly without any loosening or breakage of screws.Conclusion 3D image acquisition by C-arm combined with orthopedic robot-assisted iliosacral screw fixation through S3 vertebra is feasible,safe and effective.