Cause analysis of joint limit interlock and parameter adjustment of treatment couch in CyberKnife
10.19745/j.1003-8868.2025191
- VernacularTitle:射波刀关节限制联锁原因分析及解决方法研究
- Author:
Jun-jie LI
1
;
Rui ZHAO
;
Jin LI
;
Xiao-lei CUI
Author Information
1. 联勤保障部队第940医院放射治疗科,兰州 730050
- Publication Type:Journal Article
- Keywords:
CyberKnife;
joint limit interlock;
treatment couch parameter;
restriction node;
radiotherapy
- From:
Chinese Medical Equipment Journal
2025;46(10):114-117
- CountryChina
- Language:Chinese
-
Abstract:
The causes for CyberKnife joint limit interlock were analyzed,and the effects of the treatment couch parameters on the position of the linear accelerator were described.A method for adjusting treatment couch parameters was proposed when joint limit interlock occurred at five common restriction nodes(nodes 43,72,79,94 and 97).References were provided for solving the problem of CyberKnife joint limit interlock and ensuring the implementation of the treatment plan.[Chinese Medical Equipment Journal,2025,46(10):114-117]