TPS Modeling Validation of the First Mevion Pencil Beam Scanning Proton Therapy System in China
10.3870/j.issn.1672-0741.24.08.037
- VernacularTitle:国内首台迈胜质子治疗系统的TPS建模验证
- Author:
Weiqing WU
;
Xiaoguang LU
;
Renchao ZHENG
;
Fei LIU
;
Guangyuan HU
;
Xianglin YUAN
- Publication Type:Journal Article
- Keywords:
Penscan;
Proton treatment system;
Treatment plan system;
Modelling;
Authentication
- From:
Acta Medicinae Universitatis Scientiae et Technologiae Huazhong
2025;54(2):234-239
- CountryChina
- Language:Chinese
-
Abstract:
Projective describes Methods and Researchs, which were modeled for the first successful Proton Treatment System in the country. Methods first designed the SOBP field in water (spread-out Braggpeak, extended Peak) based on the TG-185 report, then designed the treatment plan in water based on the TT-350 report, and finally designed the treatment plan on the patient's CT image, measuring the level doses at different depths of each field in the water using the MatriXXONE ion chamber matrix, measuring Results compared with the planned system to calculate Results with a difference of 2 per cent/2 mm to validate the accuracy of the output dose. Results, with the exception of the drop zone, the central point agent of the SOBP field is less than 3% deviation and the pass rate of gamma analysis at the flat dose is more than 90%. Results in the field of treatment meet the tolerance requirements and do not use self-adapted rasters or make the field output dose calculated using static self-adapted rasters more accurate than using dynamic self-adapted rasters. The lower the beam energy, the greater the error in the dosage calculation. Concrusion, in accordance with the accuracy requirements of the planned system of calculation and treatment, successfully modelled the Proton Treatment System of the Chinese University of Science and Technology (CUTT) in conjunction with the Ziji Medical College.