- VernacularTitle:医用超声雾化器最大雾化率测试方法研究
- Author:
Xing-cai YUAN
1
;
Tao LI
1
;
Shao-wu SHI
1
;
Peng WANG
1
;
Yuan LIAO
1
;
Zheng FAN
1
Author Information
- Publication Type:Journal Article
- Keywords: medical ultrasonic nebulizer; maximum nebulization rate; volumetric method; gravimetric method; testing of nebulization rate
- From: Chinese Medical Equipment Journal 2025;46(11):84-90
- CountryChina
- Language:Chinese
- Abstract: The medical ultrasonic nebulizer was introduced in terms of composition,working principle and testing principle and method of the maximum nebulization rate.The testing characteristics of the volumetric and gravimetric methods were compared.A series of experiments were conducted using the whole-unit gravimetric approach.The initial liquid charge was indicated to be the primary factor affecting the maximum nebulization rate,and the influence increased significantly when the nebulizer's nominal volume got expanded.References were provided for medical testing staffs in selecting testing methods.