Efficacy and Safety of His-purkinje Conduction System Pacing Combined With Atrioventricular Node Ablation for the Treatment of Atrial Fibrillation Combined With Heart Failure:a Meta-analysis
10.3969/j.issn.1000-3614.2025.09.011
- VernacularTitle:希氏束-浦肯野传导系统起搏联合房室结消融治疗心房颤动合并心力衰竭的有效性及安全性Meta分析
- Author:
Jiayue HAO
1
;
Yanmin WU
1
;
Huiping LI
1
;
Pan ZHANG
1
;
Lanfang ZHANG
1
;
Zhanqi WANG
1
Author Information
1. 河北大学附属医院 心内科,保定 071000
- Publication Type:Journal Article
- Keywords:
His-Purkinje conduction system pacing;
His bundle pacing;
left bundle branch area pacing;
atrioventricular node ablation;
atrial fibrillation;
heart failure;
meta-analysis
- From:
Chinese Circulation Journal
2025;40(9):912-918
- CountryChina
- Language:Chinese
-
Abstract:
Objectives:To systemically evaluate the efficacy and safety of His-Purkinje conduction system pacing(HPCSP,including His bundle pacing[HBP]and left bundle branch area pacing[LBBAP])combined with atrioventricular node ablation for the treatment of atrial fibrillation combined with heart failure.Methods:The PubMed,Cochrane Library,Web of Science,Embase,China National Knowledge Infrastructure(CNKI),Wanfang Data,VIP,and Yiigle were searched for studies on HPCSP combined with atrioventricular node ablation for the treatment of atrial fibrillation combined with heart failure since the established until July 31,2024.Meta-analysis was performed using RevMan 5.4 and Stata 15.1.Results:A total of 13 studies were included with 1 071 patients,the success rate of HPCSP combined with atrioventricular node ablation was 93.1%.The meta-analysis results revealed that,in terms of effectiveness outcomes,compared with baseline,left ventricular end-diastolic diameter(mean difference[MD]=-3.11,95%CI:-4.16 to-2.06,P<0.000 01)was significantly reduced,New York Heart Association(NYHA)cardiac function classification(MD=-1.36,95%CI:-1.48 to-1.24,P<0.000 01)was significantly decreased,and left ventricular ejection fraction(MD=9.86,95%CI:7.02 to 12.69,P<0.000 01)was significantly increased during follow-up.The pacing QRS duration was prolonged from baseline after atrioventricular node ablation(MD=7.83,95%CI:2.79 to 12.87,P=0.002);In terms of safety outcomes,HPCSP pacing thresholds remained stable(MD=0.07,95%CI:-0.01 to 0.15,P=0.11)and impedance was decreased(MD=-78.84,95%CI:-120.21 to-37.47,P=0.000 2)during operation and follow-up.The complication rate was 7.9%,the heart failure rehospitalization rate was 4.5%and the mortality rate was 5.8%.Compared with biventricular pacing(BVP),HPCSP was correlated with shorter pacing QRS duration(MD=-39.08,95%CI:-62.35 to-15.80,P=0.001)and higher left ventricular ejection fraction(MD=4.38,95%CI:0.37 to 8.40,P=0.03),there was no significant difference in left ventricular end-diastolic diameter(MD=-9.11,95%CI:-19.93 to 1.72,P=0.100).During follow-up,LBBAP exhibited a lower pacing threshold than HBP(MD=0.61,95%CI:0.23 to 1.00,P=0.002);Endpoint event rates were similar between HBP and LBBP(RR=1.47,95%CI:0.83 to 2.60,P=0.190).Conclusions:Results of this meta-analysis demonstrate that HPCSP combined with atrioventricular node ablation for the treatment of atrial fibrillation combined with heart failure is effective and safe.HPCSP is associated with better ventricular electro-mechanical synchronization and cardiac function compared with BVP,and LBBAP pacing parameters are superior to HBP.