Finite Element Analysis on Wear of Hip Prosthesis Considering Creep under Different Physiological Activities
10.16156/j.1004-7220.2024.06.013
- VernacularTitle:考虑蠕变的髋关节假体不同生理活动下的磨损有限元分析
- Author:
Hao LI
1
;
Bin LIU
1
;
Qiwei JIAN
1
Author Information
1. 上海工程技术大学 机械与汽车工程学院,上海 201620
- Publication Type:Journal Article
- Keywords:
artificial hip joint;
wear;
creep;
physiological activities;
finite element analysis
- From:
Journal of Medical Biomechanics
2024;39(6):1099-1106
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the creep effect of ultrahigh-molecular-weight polyethylene(UHMWPE)liners on the wear of hip prostheses under different physiological activities.Methods A finite element model of a cobalt-chrome alloy(CoCr)head and UHMWPE liner was established and validated.The effects of the ISO 14242 standard gait and daily activities(standing up/sitting down,climbing stairs,and knee bending)on prosthesis wear and creep were compared and analyzed.Results Under the ISO 14242 standard gait,the maximum contact pressure of the liner decreased from 18.3 MPa to 12.2 MPa when creep was not considered,and from 15.3 MPa to 14.2 MPa when creep was considered.The maximum linear penetration depths owing to wear and creep were 0.47 mm and 0.11 mm,respectively,with a wear volume of 205 mm3.The wear under different physiological activities was ranked as follows:climbing stairs>standing up/sitting down>ISO standard gait>knee flexion.The creep was ranked as follows:ISO standard gait>stair climbing>knee bending>standing up/sitting.Conclusions The wear and creep of the acetabular liner differed significantly under different physiological activities.The maximum contact pressure of the UHMWPE liner was reduced by the creep effect.This,in turn,decreased the wear volume.Both creep and wear should be considered in the design of hip prostheses to optimize their performance and extend their service life.