Measurement method based on EBT3 film technology for quality control detection of INTRABEAM PRS500 intraoperative radiotherapy equipment
10.19745/j.1003-8868.2025105
- VernacularTitle:基于EBT3胶片技术的INTRABEAM PRS500术中放疗设备质控检测方法研究
- Author:
Yi-kun LI
1
;
Wei DING
;
Xia-yu HANG
;
Jun HU
;
Xiang-dong SUN
;
Ai-jun XU
Author Information
1. 南京中医药大学卫生经济管理学院,南京 210023;东部战区总医院秦淮医疗区放射治疗科,南京 210002
- Publication Type:Journal Article
- Keywords:
INTRABEAM PRS500 intraoperative radiotherapy equipment;
EBT3 film;
quality control testing;
dosimetric verification;
isotropy
- From:
Chinese Medical Equipment Journal
2025;46(6):47-53
- CountryChina
- Language:Chinese
-
Abstract:
Objective To propose an EBT3 film technology-based quality control measurement method for the INTRABEAM PRS500 intraoperative radiotherapy equipment to solve the problems of the traditional methods in cumbersome operation and setup error.Methods According to HJ 1198-2021 Requirements of radiation safety and protection for radiotherapy and GBZ 121-2020 Requirements for radiological protection in radiotherapy,the environmental radiation testing of the INTRABEAM PRS500 intraoperative radiotherapy equipment was carried out point by point by means of the radiation inspection instrument.The INTRABEAM PRS500 intraoperative radiotherapy equipment was characterized by a point X-ray source(XRS),and the XRS was detected in terms of the probe linearity,radiation dose,dynamic deviation,isotropy and dose rate.The EBT3 film technology was used to verify the symmetry and isotropy of the XRS planar dose of INTRABEAM PRS500 intraoperative radiotherapy equipment.Results The X-γ dose equivalent rate of each monitoring site of the INTRABEAM PRS500 intraoperative radiotherapy device was lower than the method detection limit(MDL).The results of SQA quality control showed that the INTRABEAM PRS500 intraoperative radiotherapy equipment XRS met the quality control requirements in terms of the probe linearity,radiation dose,dynamic deviation and etc,and the isotropy differences in the+X,-X,+Y,and-Y axis directions ranged from-1.40%to 1.79%,which were all within the allowable range of measurement tolerance(5.60%to 5.65%).The results of measuring the isotropy of the INTRABEAM PRS500 intraoperative radiotherapy equipment based on the EBT3 film technology showed that the dose distribution of the XRS in the directions of the+X,-X,+Y,and-Y axes at the same plane was well isotropic,and that the doses in the directions of the X and Y axes were symmetrically distributed,and that the maximum skewness value for the isotropy of the XRS in the XY plane was-1.581%,which met the requirements of AAPM TG61 report on the reference dosimetry of low-energy and medium-energy X-rays for radiotherapy and radiobiology of≤±5.3%.Conclusion The EBT3 film technology-based measurement method gains high simplicity and feasibility for the isotropy of the INTRABEAM PRS500 intraoperative radiotherapy equipment in the directions of the+X,-X,+Y,and-Y axes at the same planet,which realizes the dynamic monitoring of the dosimetric changes and facilitates the whole-process quality control management of the intraoperative radiotherapy equipment.[Chinese Medical Equipment Journal,2025,46(6):47-53]