Research Progress of Clinical Quality Control Phantoms for MRI Equipment
10.3969/j.issn.1005-5185.2025.06.007
- VernacularTitle:MRI系统临床质控模体研究进展
- Author:
Chengwei LI
1
;
Jiao LI
;
Hui XU
;
Tianrui ZHAO
;
Pu ZHANG
;
Peng ZHANG
;
Zhengshan HUANG
;
Jie SUN
Author Information
1. 中国计量科学研究院医学计量中心,北京 100029
- Publication Type:Journal Article
- Keywords:
Magnetic resonance imaging;
Phantom;
Quality control;
Review
- From:
Chinese Journal of Medical Imaging
2025;33(6):607-610,617
- CountryChina
- Language:Chinese
-
Abstract:
With the rapid increase of MRI systems in hospitals in China,national multi-sectoral strategies have been put forward to clarify requirements for improving image quality of MRI systems and preventing application risks in clinic.Quality control of MRI systems becomes an important task for regulators as well as hospital radiology departments.The tools used for quality control include imaging performance phantom and specialized function phantom,which can realize detection or calibration for parameters such as high contrast resolution,image uniformity and relaxation time.This article mainly reviews the research progress of the operation principles,common types and clinical applications for these two types of phantoms mentioned above.