Application of ArcherQA for independent dose verification of SRT plans for CyberKnife
10.3760/cma.j.cn113030-20250106-00008
- VernacularTitle:ArcherQA系统在射波刀SRT计划验证中的初步应用与探讨
- Author:
Xuyao YU
1
;
Yuwen WANG
;
Yang DONG
;
Daguang ZHANG
;
Yongchun SONG
;
Qiang REN
;
Xi PEI
;
Zhiyong YUAN
;
Wei WANG
;
Jianrong DAI
Author Information
1. 天津医科大学肿瘤医院放射治疗科,国家恶性肿瘤临床医学研究中心,天津市恶性肿瘤临床医学研究中心,天津市肿瘤防治重点实验室,天津 300060
- Publication Type:Journal Article
- Keywords:
Radiosurgery;
CyberKnife;
Stereotactic radiotherapy;
Monte Carlo;
Point dose;
Gamma pass rate
- From:
Chinese Journal of Radiation Oncology
2025;34(11):1139-1145
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the feasibility of using the domestic ArcherQA system for fast and simplified independent verification of CyberKnife (CK) stereotactic radiotherapy (SRT) plans.Methods:SRT plans of 57 patients treated with CK at Tianjin Medical University Cancer Institute and Hospital from August 2021 to August 2022 were retrospectively analyzed, including 15 intracranial, 30 pulmonary, and 12 abdominal tumors cases. Point-dose and planar-dose verifications were performed using an ionization chamber and radiochromic films embedded in a homogeneous phantom, and the results were compared with those calculated by the treatment planning system (TPS). The localization CT images and corresponding SRT plans were imported into the ArcherQA system for independent dose verification and analysis. The correlation between ArcherQA results and phantom measurements was analyzed, with comparisons of target mean dose differences and γ pass rates.Results:Phantom measurement results showed, the measured point-dose differences for intracranial, lung, and abdominal plans were -0.94% ± 3.22%, 1.92% ± 2.05%, and 2.12% ± 0.77%, respectively. The mean dose differences in target dose calculation between ArcherQA and TPS: intracranial in the gross tumor volume (GTV) regions were 0.34% ± 2.21%, lung tumor GTV were -2.47% ± 2.46%, and abdominal tumor GTV were 0.80% ± 2.61%, respectively. Among them, the abdominal GTV region showed the highest correlation between ArcherQA and measured results ( r=0.78). The average two-dimensional γ pass rates (2 mm/2%, threshold=10%) measured using phantom films were 95.92% ± 2.35% for intracranial, 95.70% ± 2.74% for lung, and 96.74% ± 3.41% for abdominal tumors plans, respectively. The three-dimensional ArcherQA results showed comparable γ pass rates (1 mm/2%, threshold=10%) for lung and abdominal GTV and PTV regions, with similar medians and data dispersion to film measurements. Conclusions:The ArcherQA system enables rapid and efficient independent dose verification of CK SRT plans without the need for additional hardware. The verification results show good correlation with phantom measurements, supporting its potential as an auxiliary quality assurance tool in clinical CK SRT implementation.