Methodological correlation and efficacy analysis of the detection of hepatitis B virus surface antigen in clinical laboratory
10.3969/j.issn.1000-484X.2025.07.034
- VernacularTitle:临床实验室检测乙型肝炎病毒表面抗原的方法学关联及效能分析
- Author:
Haixia WANG
1
;
Shulin ZHANG
;
Yingying ZHAO
;
Yangfan FENG
;
Weipeng DU
;
Jingyi LIANG
;
Jiawei LI
Author Information
1. 南阳市中心医院检验科,南阳 473000
- Publication Type:Journal Article
- Keywords:
Hepatitis B virus;
HBsAg;
Methodology;
Performance evaluation
- From:
Chinese Journal of Immunology
2025;41(7):1772-1776,1781
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To explore the evaluation and relationship of hepatitis B virus(HBV)surface antigen(HBsAg)in clinical laboratory in different detection systems,further scientifically and reasonably to explain the test results for serving clinical practices.Methods:During the period from June 2021 to July 2022,100 425 specimens of patients with screened,suspected and confirmed HBV infections were collected from the clinical departments(mainly infectious hepatology)of Nanyang Central Hospital.Detection methodology included quantitative(electrochemiluminescence and chemiluminescence),qualitative(gold standard),semi-quantita-tive(ELISA),and highly sensitive HBV-DNA(RT-PCR)methods,and then analyzed the strengths and weaknesses and closeness be-tween each methodology.The relationship between the two Roche HBsAg detection systems was analyzed by correlation analysis.The HBsAg efficacy analysis was validated using Cut/Off value setting and detection limit,which in turn analyzed the distribution of false-positive and false-negative reporting models.Results:Detection results for low-and medium-concentration HBsAg showed a correla-tion between the electrochemical luminescence semi-quantitative method and ELISA method.ELISA method still had advantages in terms of sensitivity and specificity when detecting HBsAg,and there were no significant differences compared to domestic and interna-tional HBsAg quantitative detection systems.Performance validation conducted in accordance with the CNAS-GL038 document showed that the minimum detection limit for HBsAg calculated using the ELISA method in this laboratory was 0.1 U/ml.When the ROC curve Cut/Off value was set to 0.105,the area under the curve was the largest(AUC=0.986).Based on Roche's semi-quantitative electroche-miluminescence detection and patient medical history,in the common reporting model for hepatitis B five-item detection using ELISA method,HBsAg false positives occurred most frequently when HBsAg was positive alone,and HBsAg occurred most frequently in the false positive range when the OD value was less than 0.5.In ELISA method for detecting HBsAg,as the OD value increased from 0.01 to 0.10,the number of false-negative results also increased.Roche Elecsys HBsAg Ⅱ Quant Ⅱ and Elecsys HBsAg Ⅱ testing systems exhibited good linearity under certain conditions,with a ratio of approximately 1/0.18.In Elecsys HBsAg Ⅱ Quant Ⅱ detection sys-tem,the test results of HBsAg samples diluted 400 times were highly consistent with the original test results with a coefficient of deter-mination R2=0.993 8.Conclusion:There was a certain relationship between various detection systems of HBsAg at a suitable concen-tration.The detection of HBsAg by ELISA can meet the needs of clinical detection.