Safety and efficacy of a new single-needle dialysis model in maintenance hemodialysis patients
10.3760/cma.j.cn441217-20240625-00637
- VernacularTitle:维持性血液透析患者应用新型改良单针透析模式的安全性和有效性
- Author:
Bin ZHAO
1
;
Lihong ZHANG
1
;
Shen ZHAN
1
;
Lifang LIU
1
;
Wei LIU
1
;
Shanshan GUO
1
;
Guanghui XIAO
1
;
Yuzhu WANG
1
Author Information
1. 北京大学第三医院海淀院区(北京市海淀医院)肾内科,北京 100080
- Publication Type:Journal Article
- Keywords:
Renal dialysis;
Arteriovenous fistula;
Renal replacement therapy;
Single-needle dialysis;
Dialysis adequacy
- From:
Chinese Journal of Nephrology
2025;41(2):125-127
- CountryChina
- Language:Chinese
-
Abstract:
The study was a prospective observational study. A total of 24 patients who underwent maintenance hemodialysis (MHD) at Haidian Hospital in Beijing from May 2024 to June 2024 were included as the study subjects. The safety and efficacy of a new single-needle dialysis in MHD patients were evaluated. The reasons for using single-needle dialysis included waiting for the maturity of internal fistula(7 cases, 29.17%), autogenous arteriovenous fistula thrombosis occurred (6 cases, 25.00%), puncture difficulty occurred (7 cases, 29.17%), and pain sensitivity or elderly (4 cases, 16.67%). The results showed that the average blood flow was (155.65±5.90) ml/min, total blood volume was (35.92±2.65) L during single-needle dialysis. One patient had slight puncture leakage, and the puncture success rate was 95.83%. Relevant indicators of dialysis adequacy showed that the average urea clearance (Kt/V) was 0.90±0.42, urea reduction ratio was 58.31%±7.93%, and online real-time Kt/V monitoring average value was 0.98±0.55. The results suggest that the application of the new improved single-needle dialysis mode in MHD patients is safe and effective.