Analysis on the clinical application effects and development of intelligent control system for airbag pressure of tracheal catheter
10.3969/j.issn.1672-8270.2025.07.017
- VernacularTitle:智能气管导管气囊压力控制系统研发及其临床应用效果分析
- Author:
Guangchao YUAN
1
;
Siyuan HA
;
Fang WANG
;
Weiyi ZHANG
;
Wei XIA
;
Junyan TAN
;
Yijuan DING
;
Luyan HUANG
;
Min LEI
Author Information
1. 武汉市中医医院麻醉科 武汉 430000
- Publication Type:Journal Article
- Keywords:
Intelligent control system for airbag pressure of tracheal catheter;
Airway management;
Cuff pressure;
Tracheal intubation
- From:
China Medical Equipment
2025;22(7):87-91
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To analyze the research and development of control system for airbag pressure of intelligent tracheal catheter,and its application effect in clinical practice.Methods:The hospital designed an intelligent self-adaptive control system for cuff pressure of tracheal catheter,which can rationally apply main components such as single chip microcomputer,display screen,control panel,pressure monitoring module,inflation device and solenoid valve,etc..This control system connected the measuring module of pressure through single chip microcomputer,so as to continuously detect the airbag pressure of tracheal catheter.An inflation and deflation algorithm was independently researched and developed to control air pump or solenoid valve to inject or release part of the gas to the airbag,so as to achieve intelligent control for the pressure of airbag.Sixty patients underwent endotracheal intubation who admitted to the department of general surgery of Wuhan Hospital of Traditional Chinese Medicine from January 2022 to January 2023 were selected as the study objects,and they were divided into control group(n=30)and observation group(n=30)according to random number table method.The cuff pressure of endotracheal intubation of patients in control group were managed by conventional method.The intelligent adaptive control system was adopted to manage the cuff pressure of endotracheal intubation for patients in the observation group.The complications,compliance rate of airbag pressure,the instantaneous maximum value of airbag pressure during sputum aspiration,insertion of gastric tube,and turning over of body were compared between the two groups after intubation.Results:The incidence of complication included cough,trachyphonia,dysphagia and bloody sputum after tracheal intubation in the observation group were 6.66%,which was significantly lower than 33.33%in the control group,and the difference was statistically significant(x2=6.667,P<0.05).The compliance rate of the pressure(25-30cmH2O)of airbag in the control group was 92.83%,which was significantly lower than 97.78%in the observation group,and the difference was statistically significant(x2=14.698,P<0.05).The instantaneous maximum values of airbag pressure during sputum aspiration,insertion of gastric tube and turning over were(28.24±3.65)cmH2O,(27.98±4.25)cmH2O and(28.65±4.87)cmH2O in observation group,respectively,which were significantly lower than those in control group,and the differences were significant(t=17.930,19.208,16.485,P<0.05).Conclusion:The application of intelligent adaptive control system for cuff pressure of tracheal catheter can maintain the airbag pressure of tracheal catheter at normal level(20-30 cmH2O),and reduce complications,and improve the compliance rate of airbag pressure.