Feasibility study of photon counting CT combined with “quadruple lows” technology in head and neck vascular imaging
10.3760/cma.j.cn112149-20240919-00575
- VernacularTitle:光子计数CT单能量成像联合“四低”技术在头颈血管成像中的应用价值
- Author:
Mengyuan ZHANG
1
;
Kehui NIE
;
Dian YUAN
;
Yicun ZHANG
;
Ke QI
;
Weiting ZHANG
;
Yanbo GU
;
Yan CHEN
;
Jianbo GAO
;
Jie LIU
Author Information
1. 郑州大学第一附属医院放射科,郑州450052
- Publication Type:Journal Article
- Keywords:
Tomography, X-ray computed;
Photon-counting detector;
CT angiography;
Comparative study;
Image quality
- From:
Chinese Journal of Radiology
2025;59(8):942-948
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the feasibility of head and neck vascular imaging using photon-counting detector computed tomography (PCD-CT) combined with a “quadruple lows” protocol—characterized by low contrast media volume, low iodine concentration, low injection rate, and low radiation dose—and to compare the image quality with that obtained by energy-integrating detector CT (EID-CT).Methods:A total of 105 patients with suspected cerebrovascular disease were prospectively enrolled at the First Affiliated Hospital of Zhengzhou University between April and June 2024. Patients were randomly assigned to three groups ( n=35). Group A underwent conventional head and neck CTA using EID-CT. Group B underwent PCD-CT with a protocol involving ultra-low contrast media volume, low iodine concentration, and low injection rate. Group C underwent PCD-CT with the full “quadruple low” protocol. Objective image quality parameters—including CT attenuation, image noise (standard deviation, SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR)—were measured at the ascending aorta, common carotid artery, internal carotid artery, vertebral artery, basilar artery, posterior cerebral artery, and middle cerebral artery. Two radiologists independently rated subjective image quality using a 5-point Likert scale. Differences among groups were analyzed using one-way ANOVA and the Kruskal-Wallis test. Results:Compared to Group A [contrast volume: (42.78±6.64)ml], contrast agent volume was significantly reduced in Groups B and C[ (26.26±4.45) ml and (26.54±3.83)ml, respectively], demonstrating reductions of 39% and 38% (both P<0.01). The iodine concentration was 320 mg/ml in Groups B and C, lower than 350 mg/ml in Group A (8.5%). The injection rate was also reduced in Groups B and C [(3.39±0.61) and (3.55±0.51)ml/s, respectively] compared to Group A [(4.28±0.66) ml/s], with reductions of 21% and 17% (both P<0.01). The effective dose (ED) was similar between Groups A and B [(1.40±0.15) vs. (1.40±0.19)mSv, P>0.05], while Group C demonstrated a significantly lower ED [(0.99±0.09) mSv], with a reduction of 30% compared to Group A and 29% compared to Group B (both P<0.01).In terms of objective image quality, significant differences in image noise (SD) were observed among the three groups at the vertebral artery, internal carotid artery, posterior cerebral artery, and middle cerebral artery (all P<0.05). Groups B and C showed significantly lower SD compared to Group A ( P<0.05), with no significant difference between B and C ( P>0.05). SNR was significantly higher in Groups B and C than in Group A at multiple vascular segments (all P<0.05). CNR differed only at the internal carotid artery, where Groups B and C demonstrated superior performance compared to Group A ( P<0.05).Subjective image quality scores showed no significant difference between Groups A and C ( P>0.05), while Group B had significantly higher scores than both A and C ( P<0.05). All images were deemed diagnostically acceptable. Conclusion:Compared with conventional EID-CT, PCD-CT combined with a “quadruple lows” protocol enables substantial reductions in contrast media and radiation dose while further improving image quality in head and neck CTA.