Photon-counting detector CT with virtual monoenergetic imaging in enabling a quadruple low-dose aortic CT angiography protocol: a feasibility study
10.3760/cma.j.cn112149-20240911-00552
- VernacularTitle:光子计数CT虚拟单能量成像实现主动脉CT血管成像“四低”方案的可行性研究
- Author:
Ke QI
1
;
Kehui NIE
;
Dian YUAN
;
Yicun ZHANG
;
Mengyuan ZHANG
;
Weiting ZHANG
;
Yanbo GU
;
Yan CHEN
;
Jianbo GAO
;
Jie LIU
Author Information
1. 郑州大学第一附属医院放射科,郑州 450052
- Publication Type:Journal Article
- Keywords:
Tomography, X-ray computed;
Photon-counting detector;
Angiography;
Program evaluation;
Image quality
- From:
Chinese Journal of Radiology
2025;59(10):1133-1141
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To assess the viability of reducing radiation dose, contrast media volume, injection flow rate and contrast medium concentration (quadruple low-dose protocol) by utilizing virtual monoenergetic images (VMI) in photon-counting detector CT (PCD-CT) for aortic CT angiography (CTA), while maintaining image quality in comparison to images obtained from energy-integrating detector CT (EID-CT).Methods:From April 2024 to June 2024, a total of 40 participants who underwent aortic CTA on PCD-CT were prospectively enrolled in the experimental group (PCD-CT group), while 40 patients with similar baseline characteristics who had previously undergone aortic CTA using EID-CT were retrospectively selected for the conventional group (EID-CT group). The EID-CT group used a tube voltage of 90 kVp, a contrast media volume of 60 ml of contrast, an injection flow rate of 3 ml/s, and a contrast concentration of 350 mgI/ml; the PCD-CT group used the QuantumPlus mode, with a tube voltage of 140 kVp, a total amount of iodine in the contrast media of 140 mgI/kg, and an injection flow rate=contrast media volume/(delay time+scan time), and a contrast media concentration of 320 mgI/ml. VMIs in PCD-CT group were reconstructed in 5-keV intervals ranging from 45 to 65 keV. The effective radiation dose and contrast injection protocols were recorded and compared between two groups. Objective image quality assessment was performed for each group. CT attenuation, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured at five anatomical locations (ascending aorta, aortic arch, descending aorta, abdominal aorta, and right common iliac artery), and image noise was recorded. Subjective image quality was independently evaluated by two readers using a 5-point Likert scale in a blinded manner. Based on data normality, the one-way ANOVA or Kruskal-Wallis test was used for image quality assessment, with Bonferroni-corrected post-hoc analysis for multiple comparisons.Results:There were no significant differences in the baseline characteristics between two groups (all P0.05). The PCD-CT group demonstrated significantly lower effective radiation dose [(3.88±0.65) mSv vs. (5.97±1.15)mSv], contrast media volume [(29.25±4.56) ml vs. 60 ml], and injection rate [(2.65±0.42) ml/s vs. 3 ml/s] than the EID-CT group, with reductions of 35%, 51%, and 12%, respectively (all P0.001). For objective image quality, except for the ascending aortic CT attenuation, the CT attenuation, SNR, and CNR of other vessels in the 55 keV PCD-CT group were comparable to those in the EID-CT group. Additionally, the difference in image noise between these two groups was not statistically significant ( P0.05). Concerning subjective image quality, at 55 keV, the PCD-CT group had similar image noise scores and vessel attenuation scores (both P0.05) and better visualization of renal artery branching ( P=0.001) compared to the EID-CT group. Conclusion:In comparison to EID-CT, the use of a 55 keV image in PCD-CT for aortic CTA has demonstrated reductions in radiation dose, contrast media volume, injection flow rate and contrast medium concentration, while maintaining image quality.