- VernacularTitle:基于夏克-哈特曼波前传感的紧凑型快速自动对焦眼底成像系统
- Author:
Zhe-Kai LIN
1
;
Long CHEN
1
;
Geng-Yong ZHENG
1
;
Jin-Tian HUANG
1
;
Jia-Xin DONG
1
;
Shang-Pan YANG
1
;
Wen-Zheng DING
1
;
Ding-An HAN
1
;
Xue-Hua WANG
1
;
Ya-Guang ZENG
1
Author Information
- Publication Type:Journal Article
- Keywords: portable fundus camera; autofocus; wavefront detection; Shack-Hartmann wavefront sensor
- From: Progress in Biochemistry and Biophysics 2026;53(4):1076-1086
- CountryChina
- Language:English
- Abstract: ObjectiveThe widespread adoption of portable fundus cameras for primary care and community screening is hindered by limitations in current autofocus(AF) technologies. Image-based methods relying on sharpness evaluation require iterative searches, resulting in slow convergence, while projection-based techniques are susceptible to optical artifacts and calibration errors. To address these challenges, this study introduces a novel AF system based on direct wavefront sensing, designed to deliver simultaneous high speed, high precision, and operational robustness within the compact form factor essential for portable ophthalmic devices. MethodsOur approach fundamentally reimagines the AF process by directly measuring the ocular wavefront aberration. We developed a custom portable fundus camera integrating a miniaturized Shack-Hartmann wavefront sensor (SHWS) into the optical path. An 850 nm laser diode projects a point source onto the retina via oblique illumination to minimize corneal reflections. Light scattered from this spot carries the eye’s refractive error through the imaging optics and is directed to the SHWS, positioned at a plane optically conjugate to the primary color CMOS imaging sensor. A microlens array within the SHWS samples the incident wavefront, generating a pattern of focal spots on a CCD. Real-time centroid analysis of these spots provides a map of local wavefront slopes. These measurements are processed through a singular value decomposition (SVD) algorithm to fit a Zernike polynomial basis set, enabling real-time reconstruction of the wavefront phase. The defocus component (S) is extracted from the second-order Zernike coefficients, providing a direct, quantitative measure of the refractive error in diopters. This value serves as a precise error signal in a closed-loop control system, which commands a voice-coil actuated focusing lens to its null position in a single, deterministic step, eliminating the need for iterative search algorithms. ResultsComprehensive evaluation demonstrated the system’s high performance. Testing on a calibrated model eye (OEMI-7) established a highly linear relationship between the computed defocus S and the focusing lens position across a ±20 Diopter (D) compensation range, achievable within a 5 mm mechanical travel. The system achieved a focusing precision of 0.08 D, corresponding to an 18-fold improvement over a conventional projection spot-size method tested under identical conditions. The total focus acquisition time, encompassing wavefront measurement, computation, and lens actuation, averaged under 0.5 s. Clinical validation with 25 human volunteers (50 eyes, refractive range -15 D to +10 D) confirmed practical efficacy. The wavefront-sensing AF succeeded in 92% of attempts with a mean time of 0.5 s, substantially outperforming a projection-based benchmark which achieved only a 32% success rate with an average time of 4.25 s. The system provided instantaneous directional guidance and maintained stability during minor ocular movements. Objective assessment of image quality, via amplitude contrast of retinal vasculature, showed consistent and significant enhancement following AF correction across the entire tested diopter range. ConclusionThis work successfully implements and validates a direct wavefront-sensing autofocus paradigm for portable fundus cameras. By directly quantifying and compensating for the optical defocus aberration, this method bypasses the fundamental limitations of image-processing and projection-based techniques, enabling rapid, precise, and deterministic diopter compensation. The developed system delivers an exceptional combination of a wide operational range (±20 D), high accuracy (0.08 D), fast convergence (0.5 s), and a compact physical footprint. This technology provides a practical and high-performance focusing solution capable of enhancing the reliability, throughput, and diagnostic utility of portable retinal imaging in large-scale screening applications. Future efforts will be directed towards system cost optimization and performance adaptation for diverse ocular conditions.