The impact of 125I seeds strand radian on the dose of different reference points
10.3969/j.issn.1008-794X.2025.12.009
- VernacularTitle:碘-125粒子链弧度对周围参考点剂量的影响
- Author:
Jinxin ZHAO
1
;
Dingxin WANG
;
Guozhang XING
;
Ke XU
;
Zezhou LIU
;
Huimin YU
;
Zeyang WANG
;
Juan WANG
;
Hongtao ZHANG
Author Information
1. 050051 河北石家庄 河北省人民医院肿瘤一科(河北省肿瘤放射性粒子植入诊疗中心/河北省放射性粒子
- Keywords:
radian;
125I seed strand;
dose;
brachytherapy
- From:
Journal of Interventional Radiology
2025;34(12):1333-1337
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the impact of different 125I seeds strand radian on the dose of different reference points around the seeds.Methods CT scan of self-developed radioactive particle radiation dose measurement phantom was performed,the scanned images were transferred to the three-dimensional treatment planning system(TPS).The target area at the middle level of the model was drawn.The target volume was delineated at 5 mm and 10 mm above and below the center of the phantom.125I seeds strand plans were designed with different radians,with a total length of 8 cm,seed spacing of 0 cm,activity of 0.8 mCi,and a total of 16 particles,with radians ranging from 30°to 170°,increasing by 10° increment.The point 5 mm vertically away from the center of the seeds strand towards the center was named A',and the point away from the center was named A.The point 10 mm vertically away from the center of the seeds strand towards the center was named B',and the point away from the center was named B.The doses at different radians were recorded,and the actual absorbed dose at 1-2 months after operation was calculated based on the particle activity decay formula.Results The doses at points A'and A were(218.3±23.1)and(201.5±16.0)Gy respectively(P=0.001).The actual absorbed doses at 1 month after operation were(65.5±6.9)and(60.5±4.8)Gy respectively(P=0.001),and the actual absorbed doses at 2 months after operation were(109.2±11.5)Gy and(100±7.9)Gy respectively(P=0.001).The doses at points B'and B were(95.9±11.0)Gy and(81.7±4.9)Gy respectively(P<0.001),and the actual absorbed doses at 1 month after operation were(28.8±3.3)Gy and(24.5±1.5)Gy respectively(P<0.001).The actual absorbed doses at 2 month after operation were(48.0±5.5)Gy and(41.0±2.4)Gy respectively(P<0.001).The doses at points A'and A gradually decreased with the increase of the radians,reaching the minimum value at 100 degrees,and then increased gradually,showing a cubic function change.The actual absorbed dose showed the same trend.The doses at points B'and B increased gradually with the increase of the radians,showing a cubic function change.Conclusion At different radians,the point doses and absorbed doses on the centrifugal side of the seeds strand are both less than those on the centripetal side.There is a cubic function relationship between the dose at the reference points and the radian of the seeds strand.