Design and application of protective restraint band used for the interventional catheter bed
10.3969/j.issn.1008-794X.2025.02.015
- VernacularTitle:介入手术导管床保护性约束带设计及应用
- Author:
Yan ZHANG
1
;
Zhexia JIN
;
Chunqiao WU
;
Funv SHEN
;
Longlong SHEN
;
Lumin CHEN
;
Yiyu ZHUANG
Author Information
1. 310016 浙江杭州 浙江大学医学院附属邵逸夫医院放射科
- Keywords:
interventional radiology;
physical restraint;
intraoperative nursing care
- From:
Journal of Interventional Radiology
2025;34(2):197-199
- CountryChina
- Language:Chinese
-
Abstract:
Objective To independently design a protective restraint band used for the interventional catheter bed and to evaluate its application effect.Methods The self-designed restraint belt consisted of a restraint belt body,first movable band and second movable band.A total of 200 patients,who received interventional operation at the Affiliated Run Run Shaw Hospital,School of Medicine of Zhejiang University of China from October 2022 to March 2023,were collected for this study.The patients were divided into the study group(n=100,using the self-designed restraint belt)and the control group(n=100,not using the self-designed restraint belt).The duration of surgical interruption due to patient's factors and the incidence of restraint belt-related adverse events were compared between the two groups.Results The duration of surgical interruption of the study group was(0.26±0.12)min,which was significantly shorter than(1.46±0.50)min in the control group,the difference was statistically significant(t=-23.570,P<0.01).The incidence of restraint belt-related adverse events in the study group and the control group was 0%and 7%respectively,the difference was statistically significant(x2=6.380,P=0.012).Conclusion The protective restraint band used for the interventional catheter bed can reduce the incidence of restraint belt-related adverse events and improve surgical efficiency.