Exploration on the acceptability of routine biochemical test results for serum samples with varying degrees of chylous high triglyceride
10.3969/j.issn.1673-4130.2025.07.003
- VernacularTitle:不同程度乳糜状高甘油三酯血清标本对常规生化指标检测结果影响的可接受性探讨
- Author:
Xi ZHANG
1
;
Qing YANG
;
Tao KANG
;
Ge LI
;
Zhiyang WANG
;
Li ZHANG
;
Xue WANG
;
Pei LI
;
Kan ZHONG
;
Yanguo TAN
Author Information
1. 首都医科大学附属复兴医院检验科,北京 100038
- Keywords:
lipemia;
high speed centrifugation;
chylous hypertriglyceridemia;
biochemical test
- From:
International Journal of Laboratory Medicine
2025;46(7):780-785
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the acceptability of routine biochemical test results for serum samples with varying degrees of chylous high triglyceride(TG).Methods Blood samples of 69 patients with different degrees of lipids were collected,including 33 patients with mild to moderate lipids(1.7 mmol/L≤TG<5.6 mmol/L)and 36 patients with severe lipids(TG≥5.6 mmol/L).Twenty-nine biochemical tests were detected before and after high-speed centrifugation.The result acceptability before high speed centrifugation of serum was compared with the results after high speed centrifugation as the gold standard[TG and total cholesterol(TC)before centrifugation].The acceptable criteria were subject to the following three conditions at the same time.Firstly,correlation coefficient(R2)was greater than or equal to 0.95.Secondly,the slope of linear re-gression equation was 1.00±0.05.Thirdly,for the same index,the number of samples whose result bias be-fore and after centrifugation was less than 1/2 total allowable error(TEa)in more than 90%of the total sam-ple numbers.Results Firstly,in the mild to moderate lipemia group,22 tests met the criteria,7 tests did not,including total protein(TP),albumin(ALB),TG,aspartate aminotransferase(AST),carbon dioxide(CO2),α-L-fucosidase(AFU),lactate dehydrogenase(LDH)(bias<10%),and the coincidence rate was 75.9%.In the severe lipemia group,12 tests met the criteria,17 tests did not,including pre-albumin(PA),AFU,γ-glu-tamyltransferase(γ-GT),LDH,AST,TC,direct bilirubin(DBIL),CO2,5'-nucleotidase(5'-NT),small and low-density lipoprotein cholesterol(sd-LDL-C),high density lipoprotein cholesterol(HDL-C),low-density lipoprotein cholesterol(LDL-C),adenosine deaminase(ADA),cystatin C(CysC),glycosylated albumin(GA),total bilirubin(TBIL)(bias>10%),the coincidence rate was 41.4%,and there was a statistically sig-nificant difference in the coincidence rate between the two groups(P<0.05).Secondly,there was no statisti-cally significant difference in the acceptability of results between continuous monitoring method and endpoint method detection methods(P>0.05).Conclusion Most test results of direct determination with mild or moderate lipemia samples are acceptable,and the bias of unacceptable tests is small(<10%),so it is recom-mended to issue a test report without further sample treatment.However,due to the large number of unacceptable tests and larger bias(>10%),severe lipemia samples should be determined after high-speed centrifugation.