Analysis of inter-institutional capability comparison results for medical CT quality control testing
10.13491/j.issn.1004-714X.2025.06.012
- VernacularTitle:医用CT质量控制检测机构间能力比对结果分析
- Author:
Xuan ZHOU
1
;
Jiawu FENG
1
;
Fei LIU
1
;
Gang WANG
1
;
Yongliang CHENG
2
;
Jingzhi SUN
1
Author Information
1. Hubei Provincial Hospital of Integrated Traditional Chinese & Western Medicine(Hubei Provincial Hospital of Occupational Diseases), Wuhan 430015, China;Hubei Provincial Clinical Research Center for Pneumoconiosis and Poisoning, Wuhan 430015, China.
2. Hubei Zhe'an testing Technology Co., Ltd, Wuhan 430015, China.
- Publication Type:OriginalArticles
- Keywords:
Radiological health;
CT quality control testing;
Inter-institutional capability comparison;
Z-score method
- From:
Chinese Journal of Radiological Health
2025;34(6):850-855
- CountryChina
- Language:Chinese
-
Abstract:
Objective To assess the technical levels of computed tomography (CT) quality control testing capabilities among radiological health technology service institutions in Hubei Province, China. Methods Ten radiological health technology service institutions in Hubei Province were organized to participate in an inter-institutional capability comparison of selected indicators in CT quality control testing with reference to the Specification for testing of quality control in X-ray computed tomography (WS 519—2019). The robust Z-score method was employed to evaluate the results of the capability comparison data from the ten institutions. Results Six institutions achieved a "satisfactory" rating in all assessed indicators, with a satisfaction rate of 60%. Indicators rated as "unsatisfactory" or "questionable" included "positioning light accuracy", "uniformity", "noise", and "high-contrast resolution". Conclusion The CT quality control testing capabilities of some radiological health technology service institutions in Hubei Province require further enhancement. Improved training and education for testing personnel are recommended.