Design and application of a device to prevent facial pressure injury in prone patients.
10.3760/cma.j.cn121430-20241209-01011
- Author:
Ye TIAN
1
;
Yuan YIN
1
;
Xuan JIN
2
Author Information
1. Department of Nursing, Dalian Friendship Hospital, Dalian 116001, China.
2. Ward 2, Department of Intensive Care Unit, Dalian Friendship Hospital, Dalian 116001, China. Corresponding author: Tian Ye, Email: 9804998@qq.com.
- Publication Type:Journal Article
- MeSH:
Humans;
Pressure Ulcer/prevention & control*;
Prone Position;
Equipment Design;
Facial Injuries/prevention & control*;
Respiration, Artificial/instrumentation*;
Respiratory Distress Syndrome/therapy*
- From:
Chinese Critical Care Medicine
2025;37(10):968-970
- CountryChina
- Language:Chinese
-
Abstract:
Prone position ventilation (PPV) has been widely used in the treatment strategy of patients with acute respiratory distress syndrome (ARDS). Patients undergoing PPV may develop facial edema and are at risk for pressure injuries due to prolonged prone positioning. In clinical practice, preventive measures such as repositioning, protective dressings, and pressure-relief cushions are commonly used to prevent pressure injuries. However, factors such as improper endotracheal tube placement, self-paid dressings, and delayed clearance of oral and nasal secretions have reduced the effectiveness of preventing facial pressure injuries. To address the above issues, a device for preventing pressure injuries on the faces of patients in the prone position was designed by healthcare workers in the nursing department of Dalian Friendship Hospital, and a National Utility Model Patent of China was obtained (ZL 2024 2 0340439.8). The device consists of a support plate and a circuit control system. The support plate is equipped with two support members. Support member 1 is directly fixed to the support plate, while support member 2 is connected to the support plate via a slide and a spiral rod, serving to support the patient's face and allowing for adjustment of the appropriate width according to the size of the patient's face. Inside the two support members, there are several telescopic rods, with the upper ends designed as spherical supports. The height and position of the telescopic components can be adjusted through a circuit control system, regularly changing the pressure distribution on the patient's face, thereby achieving the purpose of changing the pressure points on the face. The inner wall of support member 2 is equipped with a camera, allowing direct observation of the patient's facial condition through a monitor, avoiding compression of the eyes and nose, and promptly removing secretions from the mouth to keep the face clean, thereby reducing the risk of facial pressure-related injuries. The center of the two support members features a hollow slot, facilitating the placement of a tracheal tube. The circuit control system includes a random module, a time setting module, a control module, and a drive module. Parameters can be set as needed. When the shortest set time is reached, the random module and time setting module send instructions to the control module. Upon receiving the instructions from the time setting module and the random number from the random module, the control module transmits information to the drive module. The drive module, upon receiving the information, controls multiple telescopic rods to adjust their height and position, thereby changing the support points on the patient's face. The device features a simple structure and convenient operation, allowing for flexible adaptation to the patient's facial shape. It can be replaced with the patient's facial pressure area, providing an intuitive view of the patient's facial pressure situation. With automation and high safety, it helps reduce the risk of pressure-related injuries and lightens the workload of medical staff.