Reasons and clinical outcomes of intraocular lens exchange
10.3980/j.issn.1672-5123.2025.12.27
- VernacularTitle:人工晶状体置换术的原因及临床效果
- Author:
Yilin HAN
1
;
Xintong LI
1
;
Yifan YANG
1
;
Jingwen ZHANG
1
;
Zibo WANG
1
;
Jiaqi SHI
1
;
Fengyan ZHANG
1
Author Information
1. Department of Ophthalmology, the First Affiliated Hospital of Zhengzhou University, Zhengzhou 450000, Henan Province, China
- Publication Type:Journal Article
- Keywords:
intraocular lens exchange;
dislocation;
refractive error
- From:
International Eye Science
2025;25(12):2043-2047
- CountryChina
- Language:Chinese
-
Abstract:
AIM: To analyze the reasons and clinical outcomes of intraocular lens(IOL)exchange.METHODS:This retrospective case series study included 47 patients(53 eyes)who underwent IOL exchange surgery at the Department of Ophthalmology, the First Affiliated Hospital of Zhengzhou University, between April 2020 and May 2024, aged from 5 to 87(44.8±3.6)years old. Patients' demographics, surgical indications, surgical techniques, preoperative and postoperative uncorrected visual acuity(UCVA)and best-corrected visual acuity(BCVA), as well as postoperative complications were recorded.RESULTS:The reasons for IOL exchange included refractive error(18 cases, 23 eyes, 43%), IOL dislocation(12 cases, 13 eyes, 25%), IOL opacification(12 cases, 12 eyes, 23%), neuroadaptation failure(3 cases, 3 eyes, 6%), and patient dissatisfaction with visual quality(2 cases, 2 eyes, 4%). The surgical techniques for IOL exchange included in-the-bag IOL fixation(16 eyes, 30%), ciliary sulcus fixation(27 eyes, 51%), and scleral suture fixation(10 eyes, 19%). There was statistical significant difference between preoperative UCVA(LogMAR)and UCVA(LogMAR)at 1 d postoperatively(1.03±0.64 vs 0.50±0.46, P<0.05), and there was statistical significant difference between preoperative BCVA(LogMAR)and BCVA(LogMAR)at 1 mo postoperatively(0.41±0.37 vs 0.17±0.21, P<0.05). Postoperative complications included posterior capsule opacification in 2 eyes and IOL dislocation in 1 eye.CONCLUSION:Refractive error, IOL dislocation, and IOL opacification were the three most common reasons for IOL exchange. Although less frequent, factors such as neuroadaptation issues associated with multifocal IOLs and patient-reported visual quality dissatisfaction reflect growing expectations for improved visual outcomes. IOL exchange surgery, though technically challenging, demonstrates favorable clinical efficacy and a low complication rate, representing an effective intervention for managing postoperative IOL-related issues following cataract surgery.