Annual Report of the Korean External Quality Assessment Service on General Transfusion Medicine and General Transfusion Antibody Tests (2024)
10.15263/jlmqa.2025.47.1.1
- Author:
Han Joo KIM
1
;
Hyungsuk KIM
;
Duck CHO
;
Dae-Hyun KO
Author Information
1. Department of Laboratory Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea
- Publication Type:Report On Proficiency Testing
- From:Journal of Laboratory Medicine and Quality Assurance
2025;47(1):1-5
- CountryRepublic of Korea
- Language:English
-
Abstract:
This report provides a summary of the 2024 survey results on the external quality assessment (EQA) scheme for the general transfusion medicine test and the general transfusion antibody test programs in Korea. Proficiency testing materials were prepared at the Asan Medical Center for bi-annual distribution to participating laboratories. The accuracy rates and number of participating laboratories for the bi-annual EQAs were: ABO typing, 99.6%–99.9% (n=944, n=945); RhD typing, 99.9%–100.0% (n=929, n=930);crossmatching, 95.0%–99.2% (n=825, n=825); unexpected antibody scre ening, 99.5%–100.0% (n=363, n=367); direct antiglobulin test (DAT) using a polyspecific reagent, 99.3%–100.0% (n=296, n=299); DAT using an antiimmunoglobulin G monospecific reagent, 100.0% (n=74, n=72); and DAT using an anti-C3d monospecific reagent, 98.6%–100.0% (n=72, n=71). The 2024 EQA scheme for the transfusion medicine program has improved and maintained the standards of the participating laboratories.