A QCM Biosensor for Screening Arsenic(Ⅲ)Aptamers and Detecting Arsenic(Ⅲ)
10.13865/j.cnki.cjbmb.2024.07.1154
- VernacularTitle:一种用于筛选砷(Ⅲ)适配体和检测砷(Ⅲ)的QCM生物传感器
- Author:
Chu-Jun ZHENG
1
;
Shi-Quan QIAN
;
Xin-Pei LI
;
Xu YAN
;
Hai-Xuan HUANG
;
Yu-Xuan WANG
;
Yu-Wei YE
;
Min YUAN
Author Information
1. 上海理工大学健康科学与工程学院,上海食品快速检测工程技术研究中心,上海 200093
- Keywords:
arsenic(Ⅲ);
aptamer;
quartz crystal microbalance(QCM)-systematic evolution of ligands by exponential enrichment(SELEX)(QCM-SELEX);
sequencing
- From:
Chinese Journal of Biochemistry and Molecular Biology
2024;40(9):1282-1288
- CountryChina
- Language:Chinese
-
Abstract:
A quartz crystal microbalance(QCM)-systematic evolution of ligands by the exponential en-richment(SELEX)technique was developed to screen out aptamers with high affinity for arsenic(Ⅲ).A random single strand DNA library was designed and fixed on the mercaptoethylamine-modified crystal plate with arsenic(Ⅲ)as the target,and the free aptamer was captured in the solution,and the QCM-SELEX screening method was constructed.After 6 rounds of screening,the secondary library was se-quenced with high throughput method,and the 6S1 dissociation coefficient Kd value was 0.36 μmol/L based on QCM resonance frequency.Using 6S1 as a probe,the QCM biosensor was constructed for the detection of arsenic(Ⅲ).The sensor has a good linear relationship in the range of 0.01 μmol/L~0.2μmol/L,and the detection limit of arsenic(Ⅲ)is 5.2 nmol/L(3σ),indicatinggood selectivity.