Percutaneous closure of patent foramen ovale in a low-level position using Amplatzer ADO Ⅱ occluder:a case report
10.3969/j.issn.1004-8812.2024.06.010
- VernacularTitle:应用ADO Ⅱ封堵器经皮介入治疗低位卵圆孔未闭1例
- Author:
Hai-Bo HU
1
;
Hao-Jia HUANG
;
Zheng-Wei LI
;
Jian-Hua LÜ
;
Xiang-Bin PAN
Author Information
1. 国家心血管病中心 北京协和医学院 中国医学科学院阜外医院结构性心脏病中心,北京 100037
- Keywords:
Patent foramen ovale;
Mitral regurgitation;
Transcatheter
- From:
Chinese Journal of Interventional Cardiology
2024;32(6):346-348
- CountryChina
- Language:Chinese
-
Abstract:
Low-level patent foramen ovale nonocclusion(PFO)is a rare type of PFO in which the PFO opening is low during transcatheter closure of PFO and the distance between the PFO left atrial opening and the root of the septal side of the mitral valve is less than 9 mm,and the smallest model of the current double-disk PFO occluder(18/18)commonly used in clinical practice for low-level PFOs can touch the mitral valve,resulting in increased risk of mitral regurgitation or leaflet abrasion.The risk of mitral regurgitation or leaflet abrasion is increased,and transcatheter closure of PFO procedure can only be abandoned when encountered intraoperatively.In this article,we present a case of successful transcatheter closure of a low-level PFO using the Amplatzer ADOⅡ occluder,which provides new ideas and strategies to deel wtih this rare type of PFO.