Electromagnetic compatibility evaluation of brain-computer interface rehabilitation robots and improvement suggestions
10.19745/j.1003-8868.2024179
- VernacularTitle:脑机接口康复机器人电磁兼容性评价及整改对策
- Author:
Ming-Jian ZHU
1
;
Meng LIU
;
Mei WANG
;
Shu YANG
;
Shu-Yi ZHANG
;
Qing-Yu LI
;
Cheng-Ke ZHU
Author Information
1. 山东省医疗器械和药品包装检验研究院,济南 250101;国家药品监督管理局医用卫生材料及生物防护器械质量评价重点实验室,济南 250101
- Keywords:
brain-computer interface rehabilitation robot;
electromagnetic compatibility;
rehabilitation training;
medical electrical equipment
- From:
Chinese Medical Equipment Journal
2024;45(9):84-88
- CountryChina
- Language:Chinese
-
Abstract:
The determinations of the operation mode and basic performance conformity for evaluating the electromagnetic compatibility of brain-computer interface rehabilitation robots were described based on relevant standards of the electromag-netic compatibility of medical electrical equipment.The performance deviations and their causes during the electromagnetic compatibility evaluation of brain-computer interface rehabilitation robots were analyzed with case studies,and some improve-ment suggestions were proposed accordingly.Technical references were provided for the manufacturers of brain-computer interface rehabilitation robots.[Chinese Medical Equipment Journal,2024,45(9):84-88]