Medical staff′s exposure dose in 1 25 I implanted wards
10.3760/cma.j.issn.1674-2907.2016.13.021
- VernacularTitle:125I粒子植入病房医护人员受照剂量研究
- Author:
Jingjing HE
1
;
Hongyan SHAO
;
Yue LI
;
Hongzhen ZHOU
Author Information
1. 510515 广州,南方医科大学南方医院护理部
- Keywords:
125I;
Medical staff;
Exposure dose
- From:
Chinese Journal of Modern Nursing
2016;22(13):1857-1859
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the exposure dose of medical staff in 1 25 I implanted wards and to investigate the management measures for protection.Methods Regular monitoring was conducted using thermoluminescent dosemeter (TLD)for medical staff in 1 25 I implanted ward and ward without 1 25 I in our center. Results The radiation exposure dose to the medical staff in 1 25 I implanted ward was higher than staff in the ward without 1 25 I (t =7.42,P <0.01 ),with both below the dose limited by the international basic safety standard.The radiation exposure dose to nurses was higher than to doctors in the same ward (t =-3.74,P <0.01 ).Conclusions There are radiation to the medical staff in 1 25 I implantation ward,but the exposure dose is in the dose limited by the international basic safety standard range.The management measures for radiation protection should be taken,especially to the radiation exposure nurses.