Development of ECG Common Mode Rejection Ratio Automatic Test System
10.12455/j.issn.1671-7104.230697
- VernacularTitle:心电共模抑制比自动化测试系统的研制
- Author:
Yongqiang CHEN
1
;
Binbin YAN
;
Jun ZHANG
Author Information
1. 深圳市理邦精密仪器股份有限公司,深圳市,518122
- Keywords:
electroencephalogram;
common mode rejection ratio;
polarization voltage;
noise level;
automatic
- From:
Chinese Journal of Medical Instrumentation
2024;48(4):440-444,450
- CountryChina
- Language:Chinese
-
Abstract:
To comprehensively meet the test requirements for the common mode rejection ratio(CMRR)across different ECG-particular standards,this paper presents the design of an ECG CMRR automatic test system.The hardware component primarily consists of a test signal generation module,an automatic control network(which includes a resistance-capacitance network control module and a polarization voltage control module),and a noise level switching module.The software portion enables automatic control and user interaction.Experimental results indicate that the system is stable,reliable,and highly automated,capable of satisfying the test requirements of various ECG-particular standards,thus demonstrating a promising application prospect.