Analysis of 47 Cases of Adverse Events of Ozonic Therapeutic Instrument
10.3969/j.issn.1671-7104.2016.03.016
- VernacularTitle:47例臭氧治疗仪不良事件分析
- Author:
Ying ZHAO
1
;
Pu HUANG
;
Hui HUANG
;
Li ZHAO
;
Siwei ZHENG
;
Chen SHI
Author Information
1. 华中科技大学同济医学院附属协和医院药剂科
- Keywords:
ozonic;
intervertebral disc protrusion;
medical device;
adverse events
- From:
Chinese Journal of Medical Instrumentation
2016;40(3):214-216
- CountryChina
- Language:Chinese
-
Abstract:
With retrospective case study, this paper classified, summerized and evaluated 47 reports induced by ozonic therapeutic instrument. The results showed:25 cases of adverse events were clearly related to the use of instruments;the cases mainly reported by equipment use enterprises; the delay treatment by instrument failure was the most common adverse event performances;adverse events might be relevant to the risk factors of product quality, technical operation, patient specificity and instrument disinfection. Monitoring should be strengthened in order to reduce the ozonic therapeutic instrument adverse events and maintenance of public safety.