Development and Application of a Micro-device for Rapid Detection of Ammonia Nitrogen in Environmental Water
10.19756/j.issn.0253-3820.231287
- VernacularTitle:环境水体中氨氮微型快速检测装置的研制及应用
- Author:
Peng WANG
1
;
Yong TIAN
;
Chuan-Yu LIU
;
Wei-Liang WANG
;
Xu-Wei CHEN
;
Yan-Feng ZHANG
;
Ming-Li CHEN
;
Jian-Hua WANG
Author Information
1. 青岛理工大学环境与市政工程学院,青岛 266520
- Keywords:
Ammonia nitrogen;
Sequential injection;
Salicylic acid spectrophotometry;
On-line detection;
Light-emitting diode
- From:
Chinese Journal of Analytical Chemistry
2024;52(2):178-186,中插1-中插3
- CountryChina
- Language:Chinese
-
Abstract:
The analysis of ammonia nitrogen in real water samples is challenging due to matrix interferences and difficulties for rapid on-site analysis.On the basis of the standard method,i.e.water quality-determination of ammonia nitrogen-salicylic acid spectrophotometry(HJ 536-2009),a simple device for online detecting ammonia nitrogen was developed using a sequential injection analysis(SIA)system in this work.The ammonia nitrogen transformation system,color reaction system,and detection system were built in compatible with the SIA system,respectively.In particular,the detection system was assembled by employing light-emitting diode as the light source,photodiode as the detector,and polyvinylchloride tube as the cuvette,thus significantly reducing the volume,energy consumption and fabricating cost of the detection system.As a result,the accurate analysis of ammonia nitrogen in complex water samples was achieved.A quantitative detection of ammonia nitrogen in water sample was obtained in 12 min,along with linear range extending to 1000 μmol/L,precisions(Relative standard deviation,RSD)of 4.3%(C=10 μmol/L,n=7)and 4.2%(C=500 μmol/L,n=7),and limit of detection(LOD)of 0.65 μmol/L(S/N=3,n=7).The results of interfering experiments showed that the detection of ammonia nitrogen by the developed device was not interfered by the common coexisting ions and components,therefore the environmental water could be directly analyzed,such as reservoir water,domestic sewage,sea water and leachate of waste landfill.The analytical results were consistent with those obtained by the environmental protection standard method(Water quality determination of ammonia nitrogen-salicylic acid spectrophotometry,HJ 536-2009).In addition,the spiking recoveries were in the range of 92.3%-98.1%,further confirming the accuracy and practicality of the developed device.