1.Analysis of the layout and shielding effectiveness of medical accelerator vaults
Yajing SONG ; Zengyun NIU ; Yongzhong MA ; Shihua TAO ; Zechen FENG
Chinese Journal of Radiological Health 2025;34(2):204-208
Objective To analyze the layout and shielding effectiveness of medical accelerator vaults, and to provide a reference for the layout, shielding design, and optimization of protection of medical accelerator vaults. Methods Four medical accelerator radiotherapy vaults were selected. The layouts of these vaults were compared with the layout requirements in the radiation therapy protection standards. For each vault, the dose rates at four points of interest outside the shielding were calculated, including the primary shielding area, secondary shielding area, maze outer wall, and lateral shielding area. These values were then compared with the actual measurements obtained using a dose rate meter. Results All four vaults were located on the ground floor of the building and included a maze, with the auxiliary rooms all placed outside the treatment rooms. However, one vault was not located at one end of the building, and in another vault, the control room was exposed to direct irradiation of the useful beam. The calculated dose rates outside the primary shielding area ranged from 0.04 μSv/h to 0.62 μSv/h, while the measured values ranged from 0.10 μSv/h to 0.66 μSv/h, with the measured values being higher than the calculated ones. The calculated dose rates outside the secondary shielding area ranged from