1. Surface morphology and surface properties of Ti and TiZr implant materials
Yaoren CHANG ; Fangfang XU ; Jie LI ; Yuanhe YOU ; Chun LIU ; Lihua YIN
Chinese Journal of Stomatology 2019;54(2):118-123
Objective:
To investigate the effects of hydrophilic treatment on the surface morphology and surface properties of pure titanium and titanium-zirconium alloy implants, and to provide reference for the studies of implant surface modification.
Methods:
The pure titanium group, the hydrophilic pure titanium group, the titanium zirconium alloy group and the hydrophilic titanium-zirconium alloy group were prepared by sandblasting and acid-etching or hydrophilic sandblasting and acid-etching, (11 specimens in each group). The surface morphology and surface properties of four types of titanium specimens were analyzed by surface contact angle meter, scanning electron microscope (SEM), optical profilometer, atomic force microscope (AFM) and Raman spectrometer.
Results:
The surface contact angles of hydrophilic pure titanium and hydrophilic titanium-zirconium alloy were 1.6°±0.3° and 1.5°±0.2°, and the surface contact angles of pure titanium and titanium-zirconium alloy were 101.4°±4.6° and 96.2°±3.0°, respectively. SEM showed that the nano-protrusions on the surface of pure titanium and titanium-zirconium alloys were less or even absent, while the nano-protrusions on the surface of hydrophilic pure titanium and hydrophilic titanium-zirconium alloys were relatively more; the nano-protrusions on the surface of hydrophilic pure titanium surface were small and dense relatively, but the nano-protrusions of the hydrophilic titanium-zirconium alloy had large diameters and were dispersed relatively. The optical profiler and AFM showed that the surface roughness of hydrophilic pure titanium and hydrophilic titanium-zirconium alloy was significantly higher than that of pure titanium and titanium-zirconium alloy (