1.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
2.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
3.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
4.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
5.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
6.Safety evaluation of new drugs of traditional Chinese medicine based on human use experience.
Zhong-Qi YANG ; Ya-Qin TANG ; Hui-Min TANG ; Yan LING ; Yan-Ping DU
China Journal of Chinese Materia Medica 2025;50(3):812-816
Because of the unclear active substances, metabolic pathways, and targets of new drugs of traditional Chinese medicine(TCM), non-clinical safety evaluation often fails to accurately locate the target organs and tissue exposed to medicinal toxicity. The human use experience(HUE) contains important safety information of TCM, while the clinical safety data in the past HUE are few and have not been effectively applied. Standardized prospective HUE studies should be carried out to collect the clinical safety data, in which appropriate physical and chemical indicators(including blood, urine, and stool routine), liver biochemical indicators, kidney biochemical indicators, and cardiovascular biochemical indicators should be selected for safety evaluation, and the detection time point and sample size should be rationally designed. Importance should be attached to the observation of symptoms and signs of adverse events/reactions in patients as well as the safety information of special groups such as the elderly, children, and pregnant women. The adverse events of TCM should be observed, judged, and treated according to the theory and the diagnosis and treatment mode of TCM. The clinical safety information about the HUE should be comprehensively collected for new drugs of TCM to make up for the lack of extrapolation of toxicological test results to humans. The unique advantages of clinical origin of new drugs of TCM should be given full play for cross-reference of the results of toxicological research and the conclusions of HUE safety evaluation. In addition, benefit-risk assessment should be conducted based on HUE, and a panoramic safety evaluation system characterized by macro and micro combination and in line with the characteristics of TCM should be established to improve the success rate in the research and development of new drugs of TCM.
Humans
;
Drugs, Chinese Herbal/adverse effects*
;
Medicine, Chinese Traditional/adverse effects*
;
Drug-Related Side Effects and Adverse Reactions
;
Female
7.Association between insulin resistance and uterine volume in girls with idiopathic central precocious puberty.
Hong-Ru ZHANG ; Ya XIAO ; Shu-Qin JIANG ; Jun SUN ; Wen-Hui SHI ; Jin-Bo LI ; Ying YANG ; Wei WANG
Chinese Journal of Contemporary Pediatrics 2025;27(4):404-409
OBJECTIVES:
To investigate the association between insulin resistance and uterine volume in girls with idiopathic central precocious puberty (ICPP).
METHODS:
A retrospective study was conducted involving 61 girls diagnosed with ICPP who visited the pediatric growth and development clinic of the Third Affiliated Hospital of Zhengzhou University between January 2022 and September 2024, designated as the ICPP group, and 61 normally developing girls as the control group. The differences in insulin resistance index (homeostasis model assessment of insulin resistance, HOMA-IR), uterine volume, and other indicators between the two groups were compared, and the relationship between insulin resistance and uterine volume in these girls was analyzed.
RESULTS:
The uterine volume and HOMA-IR level in the ICPP group were significantly higher than those in the control group (P<0.05). Correlation analysis revealed that there was a positive correlation between HOMA-IR level and uterine volume in the ICPP group (rs=0.643, P<0.001). Multiple linear regression analysis indicated that as HOMA-IR increased,uterine volume in the girls tended to increase (P<0.05).
CONCLUSIONS
There is an association between insulin resistance and uterine volume in girls with ICPP, and as HOMA-IR increases, uterine volume in the girls also increases.
Humans
;
Female
;
Insulin Resistance
;
Puberty, Precocious/metabolism*
;
Uterus/pathology*
;
Child
;
Retrospective Studies
;
Organ Size
;
Linear Models
8.Chemical consitituents and hypoglycemic activity of Sophora tonkinensis
Xiao-yun YAN ; Ling-feng QIN ; Rui ZHANG ; Ya-nan YANG ; Jin-ying TIAN ; Fei YE ; Xu ZHANG ; Pei-cheng ZHANG ; Guo-cheng WANG
Acta Pharmaceutica Sinica 2024;59(11):3135-3140
Eleven compounds were isolated from the ethyl acetate fraction of the 95% aqueous ethanol extract of the roots of
9.SERVQUAL model-based investigation of education service quality for professional master's students in clinical medicine
Lu YU ; Yongjie QIN ; Zhendong JIANG ; Yang BAI ; Ya HE ; Fazhang LAN ; Cheng ZHONG
Chinese Journal of Medical Education Research 2024;23(4):465-470
Objective:To conduct a survey of the current status of education service offered to professional master's students in clinical medicine in a grade-A tertiary hospital in Chongqing, China based on the SERVQUAL model, and to discuss the problems with education service as well as measures for improvement.Methods:Based on the SERVQUAL model, a questionnaire consisting of 30 items in 5 dimensions was designed. Each item was rated for an expected score and a perceived score. Medical professional master's students in a Chongqing grade-A tertiary hospital in the first half year of 2022 were surveyed for the expected and perceived quality of education service to calculate service quality (SQ). SPSS 26.0 was used for reliability and validity analyses and descriptive statistical analyses. The results were analyzed using a two-dimensional quadrant graph.Results:A total of 263 valid questionnaires were returned, with an effective response rate of 97.77%. For both the expectation scale and the perception scale, the Cronbach's α values were >0.8, and the Kaiser-Meyer-Olkin values were >0.7, indicating good reliability and validity. The mean expected value of education service quality was 4.28, the mean perceived value was 3.96, and the mean gap value was -0.32. The SQ scores for all the items were negative, except for "provide neat suits for work". The two-dimensional quadrant analysis showed that 6 items were located in quadrant III (low-expectation and low-perception area), and 7 items were located in quadrant IV (high-expectation and low-perception area).Conclusions:The evaluation scale based on the SERVQUAL model can effectively reflect problems in postgraduate education for medical master's students, helping develop strategies to improve the quality of education service.
10.Risk factors for bronchopulmonary dysplasia in twin preterm infants:a multicenter study
Yu-Wei FAN ; Yi-Jia ZHANG ; He-Mei WEN ; Hong YAN ; Wei SHEN ; Yue-Qin DING ; Yun-Feng LONG ; Zhi-Gang ZHANG ; Gui-Fang LI ; Hong JIANG ; Hong-Ping RAO ; Jian-Wu QIU ; Xian WEI ; Ya-Yu ZHANG ; Ji-Bin ZENG ; Chang-Liang ZHAO ; Wei-Peng XU ; Fan WANG ; Li YUAN ; Xiu-Fang YANG ; Wei LI ; Ni-Yang LIN ; Qian CHEN ; Chang-Shun XIA ; Xin-Qi ZHONG ; Qi-Liang CUI
Chinese Journal of Contemporary Pediatrics 2024;26(6):611-618
Objective To investigate the risk factors for bronchopulmonary dysplasia(BPD)in twin preterm infants with a gestational age of<34 weeks,and to provide a basis for early identification of BPD in twin preterm infants in clinical practice.Methods A retrospective analysis was performed for the twin preterm infants with a gestational age of<34 weeks who were admitted to 22 hospitals nationwide from January 2018 to December 2020.According to their conditions,they were divided into group A(both twins had BPD),group B(only one twin had BPD),and group C(neither twin had BPD).The risk factors for BPD in twin preterm infants were analyzed.Further analysis was conducted on group B to investigate the postnatal risk factors for BPD within twins.Results A total of 904 pairs of twins with a gestational age of<34 weeks were included in this study.The multivariate logistic regression analysis showed that compared with group C,birth weight discordance of>25%between the twins was an independent risk factor for BPD in one of the twins(OR=3.370,95%CI:1.500-7.568,P<0.05),and high gestational age at birth was a protective factor against BPD(P<0.05).The conditional logistic regression analysis of group B showed that small-for-gestational-age(SGA)birth was an independent risk factor for BPD in individual twins(OR=5.017,95%CI:1.040-24.190,P<0.05).Conclusions The development of BPD in twin preterm infants is associated with gestational age,birth weight discordance between the twins,and SGA birth.

Result Analysis
Print
Save
E-mail