1.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
2.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
3.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
4.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
5.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
6.International clinical practice guideline on the use of traditional Chinese medicine for functional dyspepsia (2025).
Sheng-Sheng ZHANG ; Lu-Qing ZHAO ; Xiao-Hua HOU ; Zhao-Xiang BIAN ; Jian-Hua ZHENG ; Hai-He TIAN ; Guan-Hu YANG ; Won-Sook HONG ; Yu-Ying HE ; Li LIU ; Hong SHEN ; Yan-Ping LI ; Sheng XIE ; Jin SHU ; Bin-Fang ZENG ; Jun-Xiang LI ; Zhen LIU ; Zheng-Hua XIAO ; Jing-Dong XIAO ; Pei-Yong ZHENG ; Shao-Gang HUANG ; Sheng-Liang CHEN ; Gui-Jun FEI
Journal of Integrative Medicine 2025;23(5):502-518
Functional dyspepsia (FD), characterized by persistent or recurrent dyspeptic symptoms without identifiable organic, systemic or metabolic causes, is an increasingly recognized global health issue. The objective of this guideline is to equip clinicians and nursing professionals with evidence-based strategies for the management and treatment of adult patients with FD using traditional Chinese medicine (TCM). The Guideline Development Group consulted existing TCM consensus documents on FD and convened a panel of 35 clinicians to generate initial clinical queries. To address these queries, a systematic literature search was conducted across PubMed, EMBASE, the Cochrane Library, China National Knowledge Infrastructure (CNKI), VIP Database, China Biology Medicine (SinoMed) Database, Wanfang Database, Traditional Medicine Research Data Expanded (TMRDE), and the Traditional Chinese Medical Literature Analysis and Retrieval System (TCMLARS). The evidence from the literature was critically appraised using the Grading of Recommendations Assessment, Development, and Evaluation (GRADE) approach. The strength of the recommendations was ascertained through a consensus-building process involving TCM and allopathic medicine experts, methodologists, pharmacologists, nursing specialists, and health economists, leveraging their collective expertise and empirical knowledge. The guideline comprises a total of 43 evidence-informed recommendations that span a range of clinical aspects, including the pathogenesis according to TCM, diagnostic approaches, therapeutic interventions, efficacy assessments, and prognostic considerations. Please cite this article as: Zhang SS, Zhao LQ, Hou XH, Bian ZX, Zheng JH, Tian HH, Yang GH, Hong WS, He YY, Liu L, Shen H, Li YP, Xie S, Shu J, Zeng BF, Li JX, Liu Z, Xiao ZH, Xiao JD, Zheng PY, Huang SG, Chen SL, Fei GJ. International clinical practice guideline on the use of traditional Chinese medicine for functional dyspepsia (2025). J Integr Med. 2025; 23(5):502-518.
Dyspepsia/drug therapy*
;
Humans
;
Medicine, Chinese Traditional/methods*
;
Practice Guidelines as Topic
;
Drugs, Chinese Herbal/therapeutic use*
8.Multidisciplinary expert consensus on weight management for overweight and obese children and adolescents based on healthy lifestyle
HONG Ping, MA Yuguo, TAO Fangbiao, XU Yajun, ZHANG Qian, HU Liang, WEI Gaoxia, YANG Yuexin, QIAN Junwei, HOU Xiao, ZHANG Yimin, SUN Tingting, XI Bo, DONG Xiaosheng, MA Jun, SONG Yi, WANG Haijun, HE Gang, CHEN Runsen, LIU Jingmin, HUANG Zhijian, HU Guopeng, QIAN Jinghua, BAO Ke, LI Xuemei, ZHU Dan, FENG Junpeng, SHA Mo, Chinese Association for Student Nutrition & ; Health Promotion, Key Laboratory of Sports and Physical Fitness of the Ministry of Education,〖JZ〗 Engineering Research Center of Ministry of Education for Key Core Technical Integration System and Equipment,〖JZ〗 Key Laboratory of Exercise Rehabilitation Science of the Ministry of Education
Chinese Journal of School Health 2025;46(12):1673-1680
Abstract
In recent years, the prevalence of overweight and obesity among children and adolescents has risen rapidly, posing a serious threat to their physical and mental health. To provide scientific, systematic, and standardized weight management guidance for overweight and obese children and adolescents, the study focuses on the core concept of healthy lifestyle intervention, integrates multidisciplinary expert opinions and research findings,and proposes a comprehensive multidisciplinary intervention framework covering scientific exercise intervention, precise nutrition and diet, optimized sleep management, and standardized psychological support. It calls for the establishment of a multi agent collaborative management mechanism led by the government, implemented by families, fostered by schools, initiated by individuals, optimized by communities, reinforced by healthcare, and coordinated by multiple stakeholders. Emphasizing a child and adolescent centered approach, the consensus advocates for comprehensive, multi level, and personalized guidance strategies to promote the internalization and maintenance of a healthy lifestyle. It serves as a reference and provides recommendations for the effective prevention and control of overweight and obesity, and enhancing the health level of children and adolescents.
9.A reporter gene assays for bioactivity determination of human chorinonic gonadotropin
Ying HUANG ; Xiao-ming ZHANG ; He-yang LI ; Lü-yin WANG ; Hui ZHANG ; Ping LÜ ; Jing LI ; Xiang-dong GAO ; Cheng-gang LIANG
Acta Pharmaceutica Sinica 2024;59(2):432-438
This study constructed a LHCGR-CRE-luc-HEK293 transgenic cell line according to the activation of the cAMP signaling pathway after recombinant human chorionic gonadotropin binding to the receptor. The biological activity of recombinant human chorionic gonadotropin was assayed using a luciferase assay system. The relative potency of the samples was calculated using four-parameter model. And the method conditions were optimized to validate the specificity, relative accuracy, precision and linearity of the method. The results showed that there was a quantitative potency relationship of human chorinonic gonadotropin (hCG) in the method and it was in accordance with the four-parameter curve. After optimization, the conditions were determined as hCG dilution concentration of 2.5 μg·mL-1, dilution ratio of 1∶4, cell number of 10 000-15 000 cells/well, and induction time of 6 h. The method had good specificity, relative accuracy with relative bias ranging from -8.9% to 3.4%, linear regression equation correlation coefficient of 0.996, intermediate precision geometric coefficient of variation ranging from 3.3% to 15.0%, and linearity range of 50% to 200%. This study successfully established and validated a reporter gene method to detect hCG biological activity, which can be used for hCG biological activity assay and quality control.
10.Comparison and evaluation of three thyroid imaging reporting and data systems for medullary thyroid carcinoma
Jing YU ; Yuanjing HUANG ; Xiao MA ; Yaning KUANG ; Gang DONG ; Kefei CUI
Chinese Journal of Endocrine Surgery 2024;18(4):505-509
Objective:To investigate the diagnostic performance of different thyroid imaging reporting and data systems (TI-RADS) in the diagnosis of medullary thyroid carcinoma (MTC) .Methods:A total of 160 thyroid nodules diagnosed as MTC by postoperative pathology from Aug. 2011 to Aug. 2022 at the First Affiliated Hospital of Zhengzhou University were included. Additionally, 160 papillary thyroid carcinomas (PTC) and 160 benign nodules were randomly selected as controls during the same period. Differences in gender, age, nodule diameter and various ultrasound features were observed. The nodules were classified according to American College of Radiology (ACR) TI-RADS, artificial intelligence (AI) TI-RADS and Chinese (C TI-RADS). Then, receiver operating characteristic curve (ROC) were plotted to calculate the diagnostic value. The Kendall concordance coefficient was used to evaluate the interobserver consistency of each TI-RADS system.Results:There was no statistically significant difference in gender among the three groups ( χ2=1.17, P=0.558). However, significant differences were observed in age and nodule diameter ( F=12.08,40.12, P<0.001 for both). The area under ROC (AUC) for diagnosing MTC and benign nodules using ACR, AI, and C-TIRADS were 0.762, 0.773, and 0.761, respectively, with no statistically significant differences ( Z=1.33, 0.01, 0.87, P=0.183, 0.994, 0.386). However, the sensitivity of C TI-RADS (87.5%) was lower than that of ACR and AI TI-RADS (both 95.0%) ( P=0.018). After combining the biopsy threshold, the false negative rate of C-TIRADS was lower than that of ACR (30.6% vs. 41.3%) ( P=0.048) and AI TIRADS (30.6% vs. 43.1%) ( P=0.020). The inter-observer diagnostic consistency of C-TIRADS was superior to ACR (0.884 vs. 0.819, P<0.001) and AI TIRADS (0.884 vs. 0.839) ( P<0.001) . Conclusions:AI and ACR TI-RADS have higher sensitivity in diagnosing MTC, while C TI-RADS has a lower puncture missed diagnosis rate. AI has similar diagnostic performance to ACR TI-RADS and can replace ACR TI-RADS.