1.Research on Fault Risk Identification and Control of CT Simulator Based on FMEA.
Fan BI ; Haisheng HU ; Wenyong TU
Chinese Journal of Medical Instrumentation 2019;43(4):303-306
OBJECTIVE:
To explore the high-risk fault risk of CT simulator and the main causes of the risk, and to put forward effective risk management strategies.
METHODS:
The failure mode and effect analysis method was used to identify and control the operational fault risk of CT simulator.
RESULTS:
5 major fault components, 8 fault failure models and 17 failure causes were analyzed. The top 5 failure causes are:anode target surface burn caused by direct scanning without warming up the tube (590.4), tube failure (518.2), burnout of joints caused by aging of high voltage cables (424.2), motor carbon brush wear (304.8) and belt break (296.4).
CONCLUSIONS
The failure mode and effect analysis method can effectively identify the risk of equipment failure, and thus specifically formulate risk management and control measures to ensure the normal operation of equipment and the safety of doctors and patients.
Equipment Failure
;
statistics & numerical data
;
Humans
;
Risk Management
;
methods
;
Tomography, X-Ray Computed
;
standards
2.Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus(R) in Pig Liver.
Eun Sun LEE ; Jeong Min LEE ; Kyung Won KIM ; In Joon LEE ; Joon Koo HAN ; Byung Ihn CHOI
Korean Journal of Radiology 2013;14(2):194-201
OBJECTIVE: To determine in vivo efficacy of radiofrequency ablation (RFA) in porcine liver by using 15-gauge Octopus(R) (15-G Octopus(R)) electrodes to create a large coagulation. MATERIALS AND METHODS: A total of 18 coagulations were created by using a 180-W generator and 15-G Octopus(R) electrodes during laparotomy, performed in 14 pigs. Coagulation necrosis was created in the pig livers by the use of one of three RFA protocols: 1) group A, monopolar RFA using a 15-G Octopus(R) electrode with a 5-mm inter-electrode distance (n = 4); 2) group B, monopolar RFA using a 15-G Octopus(R) electrode with a 10-mm inter-electrode distance (n = 6); and 3) group C, switching monopolar RFA using two 15-G Octopus(R) electrodes (n = 8). The energy efficiency, shape, maximum and minimum diameters (Dmx and Dmi), and the volume of the coagulation volume were measured in each group. The Summary statistics were obtained and Mann-Whitney test was were performed. RESULTS: The mean ablated volume of each group was 49.23 cm3 in A, 64.11 cm3 in B, and 72.35 cm3 in C. The mean Dmx and Dmi values were 5.68 cm and 4.58 cm in A and 5.97 cm and 4.97 cm in B, respectively. In group C, the mean diameters of Dmx and Dmi were 6.80 cm and 5.11 cm, respectively. The mean ratios of Dmi/Dmx were 1.25, 1.20, and 1.35 in groups A, B, and C, respectively. There was one animal death during the RFA procedure, the cause of which could not be subsequently determined. However, there were no other significant, procedure-related complications during the seven-hour-delayed CT scans. CONCLUSION: RFA procedures using 15-G Octopus(R) electrodes are useful and safe for creating a large ablation in a single electrode model as well as in the multiple electrodes model.
Animals
;
Catheter Ablation/*methods
;
Electrocoagulation/*instrumentation
;
*Electrodes
;
Liver/radiography/*surgery
;
Statistics, Nonparametric
;
Swine
;
Tomography, X-Ray Computed
3.Clinical Application of Modified Technique of Temporal HRCT for Dehiscence of Tympanic Segment of Facial Nerve.
Seung Chul OH ; Hee Young HWANG
Korean Journal of Otolaryngology - Head and Neck Surgery 1998;41(3):313-317
BACKGROUND AND OBJECTIVES: High Resolution Computed Tomography(HRCT) of the temporal bone is the most accurate diagnostic tool and is popularly used in the chronic otitis media with cholesteatoma. However, the bony dehiscence of tympanic segment of the facial canal may be difficult to evaluate by the conventional HRCT of the temporal bone. This study was undertaken to develop a modified technique of axial CT scan for evaluation of bony dehiscence of tympanic segment of the facial canal. MATERIALS AND METHODS: A prospective study was performed in 33 patients with chronic otitis media with cholesteatoma. These patients had also underwent canal down mastoidectomy from August of 1996 through July of 1997. A correlation study of the CT and surgical findings at 40 degree and 30 degree axial scans was carried out. RESULTS: The sensitivity and specificity of this method for dehiscence of tympanic segment of the facial canal were 88.9% and 95.6%, respectively and the diagnostic predictability was 93.9%. However, the sensitivity, specificity, and diagnostic predictability of the 30 degree axial scan were 77.8%, 83.3%, and 81.8%, respectively. CONCLUSION: We think that the modified technique of axial CT scan of temporal bone is an effective method for detecting the dehiscence of tympanic segment of facial canal.
Cholesteatoma
;
Facial Nerve*
;
Humans
;
Otitis Media
;
Prospective Studies
;
Sensitivity and Specificity
;
Statistics as Topic
;
Temporal Bone
;
Tomography, X-Ray Computed
4.Trends of CT Use in the Pediatric Emergency Department in a Tertiary Academic Hospital of Korea during 2001-2010.
Hye Yeon OH ; Eun Young KIM ; Jee Eun KIM ; Yoo Jin KIM ; Hye Young CHOI ; Jinseong CHO ; Hyuk Jun YANG ; Eell RYOO
Korean Journal of Radiology 2012;13(6):771-775
OBJECTIVE: We wanted to assess the trends of computed tomography (CT) examinations in a pediatric emergency department (ED). MATERIALS AND METHODS: We searched the medical database to identify the pediatric patients who had visited the ED, and the number of CTs conducted from January 2001 to December 2010. We analyzed the types of CTs, according to the anatomic region, and the patients who underwent CT examinations for multiple regions. Data were stratified, according to the patient age (< 13 years and 13 < or = ages < 18 years). RESULTS: The number of CTs performed per 1000 patients increased by 92% during the 10-year period (per 1000 patients, increased from 50.1 CTs in 2001 to 156.5 CTs in 2006, and then decreased to 96.0 CTs in 2010). Although head CTs were performed most often (74.6% of all CTs), facial bone CTs showed the largest rate of increase (3188%) per 1000 patients, followed by cervical CTs (642%), abdominal CTs (474%), miscellaneous CTs (236%), chest CTs (89%) and head CTs (39%). The number of patients who had CT examinations for multiple regions in the same day showed a similar pattern of increase, to that of overall CT examinations. Increase of CT utilization was more pronounced in adolescents than in pediatric patients younger than 13 years (189% vs. 59%). CONCLUSION: The utilization of CTs increases from 2001 to 2006, and has declined since 2006. The increase of CTs is more pronounced in adolescents, and facial bone CTs prevail in increased number of examination followed by cervical CTs, abdominal CTs, miscellaneous CTs, chest CTs, and head CTs.
Adolescent
;
Child
;
Child, Preschool
;
Emergency Service, Hospital/statistics & numerical data/*trends
;
Female
;
Humans
;
Male
;
Multidetector Computed Tomography
;
Republic of Korea
;
Tertiary Care Centers/statistics & numerical data/*trends
;
Tomography, X-Ray Computed/*trends/utilization
5.Reconstruction Algorithms Influence the Follow-Up Variability in the Longitudinal CT Emphysema Index Measurements.
Bruno HOCHHEGGER ; Klaus Loureiro IRION ; Edson MARCHIORI ; Jose Silva MOREIRA
Korean Journal of Radiology 2011;12(2):169-175
OBJECTIVE: We wanted to compare the variability in the longitudinal emphysema index (EI) measurements that were computed with standard and high resolution (HR) reconstruction algorithms (RAs). MATERIALS AND METHODS: We performed a retrospective review of 475 patients who underwent CT for surveillance of lung nodules. From this cohort, 50 patients (28 male) were included in the study. For these patients, the baseline and follow-up scans were acquired on the same multidetector CT scanner and using the same acquisition protocol. The CT scans were reconstructed with HR and standard RAs. We determined the difference in the EI between CT1 and CT2 for the HR and standard RAs, and we compared the variance of these differences. RESULTS: The mean of the variation of the total lung volume was 0.14 L (standard deviation [SD] = 0.13 L) for the standard RA and 0.16 L (SD = 0.15 L) for the HR RA. These differences were not significant. For the standard RA, the mean variation was 0.13% (SD = 0.44%) for EI -970 and 0.4% (SD = 0.88%) for EI -950; for the HR RA, the mean variation was 1.9% (SD = 2.2%) for EI -970 and 3.6% (SD = 3.7%) for EI -950. These differences were significant. CONCLUSION: Using an HR RA appears to increase the variability of the CT measurements of the EI.
Aged
;
Algorithms
;
Artifacts
;
Female
;
Humans
;
Imaging, Three-Dimensional
;
Male
;
Pulmonary Emphysema/*radiography
;
Radiographic Image Interpretation, Computer-Assisted/*methods
;
Retrospective Studies
;
Statistics, Nonparametric
;
*Tomography, X-Ray Computed
6.A Magnetic Resonance-based Seed Localization Method for I-125 Prostate Implants.
Rena J LEE ; Hyun Suk SUH ; Kyung Ja LEE ; Soome LIM ; Yookyung KIM ; Sungkyu KIM ; Jinho CHOI
Journal of Korean Medical Science 2007;22(Suppl):S129-S133
This study was performed to develop and evaluate a semi-automatic seed localization algorithm from magnetic resonance (MR) images for interstitial prostate brachytherapy. The computerized tomography (CT) and MR images (3 mm-slice thickness) of six patients who had received real-time MR imaging-guided interstitial prostate brachytherapy were obtained. An automatic seed localization method was performed on CT images to obtain seed coordinates, and an algorithm for seed localization from MR images of the prostate was developed and tested. The resultant seed distributions from MR images were then compared to CT-derived distribution by matching the same seeds and calculating percent volume receiving 100% of the prescribed dose and the extent of the volume in 3-dimensions. The semiautomatic seed localization method made it possible to extract more than 90% of the seeds with either less than 8% of noises or 3% of missing seeds. The mean volume difference obtained from CT and MR receiving 100% of the prescribed dose was less than 3%. The maximum extent of the volume receiving the prescribed dose were 0.3, 0.6, and 0.2 cm in x, y, and z directions, respectively. These results indicate that the algorithm is very useful in identifying seeds from MR image for post-implant dosimety.
Algorithms
;
Brachytherapy/*methods/statistics & numerical data
;
Humans
;
Iodine Radioisotopes/*administration & dosage
;
*Magnetic Resonance Imaging, Interventional/statistics & numerical data
;
Male
;
Prostatic Neoplasms/*pathology/radiography/*radiotherapy
;
Radiotherapy Dosage
;
Radiotherapy Planning, Computer-Assisted
;
Tomography, X-Ray Computed/statistics & numerical data
;
Tumor Burden
7.Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT.
Mannudeep K KALRA ; Michael M MAHER ; Srinivasa R PRASAD ; M Sikandar HAYAT ; Michael A BLAKE ; Jose VARGHESE ; Elkan F HALPERN ; Sanjay SAINI
Korean Journal of Radiology 2003;4(4):234-238
OBJECTIVE: We evaluated the association between patients' weight and abdominal cross-sectional dimensions and CT image quality. MATERIALS AND METHODS: We prospectively evaluated 39 cancer patients aged more than 65 years with multislice CT scan of abdomen. All patients underwent equilibrium phase contrast-enhanced abdominal CT with 4 slices (from top of the right kidney) obtained at standard tube current (240 280 mA). All other scanning parameters were held constant. Patients' weight was measured just prior to the study. Cross-sectional abdominal dimensions such as circumference, area, average anterior abdominal wall fat thickness and, anteroposterior and transverse diameters were measured in all patients. Two subspecialty radiologists reviewed randomized images for overall image quality of abdominal structures using 5-point scale. Non-parametric correlation analysis was performed to determine the association of image quality with patients' weight and cross-sectional abdominal dimensions. RESULTS: A statistically significant negative linear correlation of 0.46, 0.47, 0.47, 0.58, 0.56, 0.54, and 0.56 between patient weight, anterior abdominal fat thickness, anteroposterior and transverse diameter, circumference, cross-sectional area and image quality at standard scanning parameters was found (p< 0.01). CONCLUSION: There is a significant association between image quality, patients' weight and cross-sectional abdominal dimensions. Maximum transverse diameter of the abdomen has the strongest association with subjective image quality.
Aged
;
Aged, 80 and over
;
Body Weight/*physiology
;
Body Weights and Measures/*statistics & numerical data
;
Contrast Media/administration & dosage
;
Female
;
Human
;
Male
;
Observer Variation
;
Prospective Studies
;
Radiation Dosage
;
Radiography, Abdominal/*methods
;
Statistics, Nonparametric
;
Tomography, X-Ray Computed/*methods
8.Effects of Dual-Energy CT with Non-Linear Blending on Abdominal CT Angiography.
Sulan LI ; Chaoqin WANG ; Xiaochen JIANG ; Ge XU
Korean Journal of Radiology 2014;15(4):430-438
OBJECTIVE: To determine whether non-linear blending technique for arterial-phase dual-energy abdominal CT angiography (CTA) could improve image quality compared to the linear blending technique and conventional 120 kVp imaging. MATERIALS AND METHODS: This study included 118 patients who had accepted dual-energy abdominal CTA in the arterial phase. They were assigned to Sn140/80 kVp protocol (protocol A, n = 40) if body mass index (BMI) < 25 or Sn140/100 kVp protocol (protocol B, n = 41) if BMI > or = 25. Non-linear blending images and linear blending images with a weighting factor of 0.5 in each protocol were generated and compared with the conventional 120 kVp images (protocol C, n = 37). The abdominal vascular enhancements, image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and radiation dose were assessed. Statistical analysis was performed using one-way analysis of variance test, independent t test, Mann-Whitney U test, and Kruskal-Wallis test. RESULTS: Mean vascular attenuation, CNR, SNR and subjective image quality score for the non-linear blending images in each protocol were all higher compared to the corresponding linear blending images and 120 kVp images (p values ranging from < 0.001 to 0.007) except for when compared to non-linear blending images for protocol B and 120 kVp images in CNR and SNR. No significant differences were found in image noise among the three kinds of images and the same kind of images in different protocols, but the lowest radiation dose was shown in protocol A. CONCLUSION: Non-linear blending technique of dual-energy CT can improve the image quality of arterial-phase abdominal CTA, especially with the Sn140/80 kVp scanning.
Adult
;
Aged
;
Angiography/*methods
;
Body Mass Index
;
Female
;
Humans
;
Male
;
Middle Aged
;
Observer Variation
;
Radiation Dosage
;
Radiographic Image Enhancement/methods
;
Radiographic Image Interpretation, Computer-Assisted/*methods
;
Radiography, Abdominal/*methods
;
Signal-To-Noise Ratio
;
Statistics, Nonparametric
;
Tomography, X-Ray Computed/*methods
9.Value of CT in the Discrimination of Fatal from Non-Fatal Stercoral Colitis.
Cheng Hsien WU ; Chen Chih HUANG ; Li Jen WANG ; Yon Cheong WONG ; Chao Jan WANG ; Wan Chak LO ; Being Chuan LIN ; Yung Liang WAN ; Chuen HSUEH
Korean Journal of Radiology 2012;13(3):283-289
OBJECTIVE: Clinical presentation and physical signs may be unreliable in the diagnosis of stercoral colitis (SC). This study evaluates the value of computed tomography (CT) in distinguishing fatal from non-fatal SC. MATERIALS AND METHODS: Ten patients diagnosed as SC were obtained from inter-specialist conferences. Additional 13 patients with suspected SC were identified via the Radiology Information System (RIS). These patients were divided into two groups; fatal and non-fatal SCs. Their CT images are reviewed by two board-certified radiologists blinded to the clinical data and radiographic reports. RESULTS: SC occurred in older patients and displayed no gender predisposition. There was significant correlation between fatal SC and CT findings of dense mucosa (p = 0.017), perfusion defects (p = 0.026), ascites (p = 0.023), or abnormal gas (p = 0.033). The sensitivity, specificity, and accuracy of dense mucosa were 71%, 86%, and 81%, respectively. These figures were 75%, 79%, and 77% for perfusion defects; 75%, 80%, and 78% for ascites; and 50%, 93%, and 78% for abnormal gas, respectively. Each CT sign of mucosal sloughing and pericolonic abscess displayed high specificity of 100% and 93% for diagnosing fatal SC, respectively. However, this did not reach statistical significance in diagnosing fatal SC. CONCLUSION: CT appears to be valuable in discriminating fatal from non-fatal SC.
Adult
;
Aged
;
Aged, 80 and over
;
Chi-Square Distribution
;
Colitis/mortality/*radiography
;
Contrast Media/diagnostic use
;
Diagnosis, Differential
;
Fecal Impaction/mortality/*radiography
;
Female
;
Humans
;
Male
;
Middle Aged
;
Retrospective Studies
;
Risk Factors
;
Sensitivity and Specificity
;
Statistics, Nonparametric
;
Tomography, X-Ray Computed/*methods
10.Evaluation of Emphysema in Patients with Asthma Using High-resolution CT.
Ki Young HONG ; June Hyuk LEE ; Sung Woo PARK ; Jae Hak JOO ; Do Jin KIM ; Sung Heuk MOON ; Soo Taek UH ; Yong Hoon KIM ; Choon Sik PARK ; Jae Seung PARK
The Korean Journal of Internal Medicine 2002;17(1):24-30
BACKGROUND: Bronchial asthma is a clinical syndrome characterized by reversibility of airway obstruction. However, many asthmatics have evidence of residual airway obstruction. It has become evident that the repair of the chronic inflammatory process can lead to various irreversible changes. It is generally accepted that the most common cause for the change is cigarette smoking but it is controversial whether asthma progresses to emphysema. High resolution computed tomography (HRCT) is more sensitive and more accurate than chest plain films in determining the type and extent of emphysema. This study was carried out to determine whether asthma can be a cause of emphysema without the effect of cigarette smoking and to evaluate clinical characteristics in asthmatics with emphysema. METHODS: We studied 58 asthmatic patients with reversible airway obstruction and evaluated the presence of emphysema using HRCT and pulmonary function test. According to HRCT findings, they were divided into 2 groups : Asthmatics with emphysema and the ones without emphysema. REWSULTS: Of the 58 patients, 7 were revealed to have emphysema. (1) 6 asthmatics with emphysema were smokers, but one patient was a nonsmoker. (2) Highly significant differences between asthmatics with and without emphysema were found in cigarette smoking (p< 0.01) and smoking consumption (p< 0.01). (3) There were no significant differences in the duration of asthma, age or sex between patients with and without emphysema. (4) There were no significant differences in FEV1(%), FEV1/FVC (%), diffusing capacity for carbon monoxide (DLco) (%) and DLco/alveolar volume between patients with and without emphysema (5) Differences between asthma patients without emphysema and those with emphysema were found to be significant in bronchial wall thickeness (p< 0.05) and in total Ig E levels (p=0.07). CONCLUSION: These results indicate that smoking is a main factor in causing emphysema in asthmatics.
Adolescent
;
Adult
;
Aged
;
Asthma/*complications/physiopathology/radiography
;
Comparative Study
;
Female
;
Human
;
Male
;
Middle Age
;
Pulmonary Emphysema/*etiology/physiopathology/radiography
;
Respiratory Function Tests/statistics & numerical data
;
Smoking/adverse effects
;
Tomography, X-Ray Computed/methods