1.Esophageal Baseline Impedance From High-resolution Impedance Manometry Correlates With Mean Nocturnal Baseline Impedance From pH-impedance Monitoring
Anthony HORTON ; Brian SULLIVAN ; Katie CHARLES ; Thasha MCINTOSH ; Andrea DAVIS ; Ziad GELLAD ; Rahul SHIMPI ; C Prakash GYAWALI ; Amit PATEL
Journal of Neurogastroenterology and Motility 2020;26(4):455-462
Background/Aims:
Esophageal baseline impedance (BI) can be extracted from pH-impedance tracings as mean nocturnal baseline impedance (MNBI), and from high-resolution impedance manometry (HRIM), but it is unknown if values are similar between acquisition methods across HRIM manufacturers. We aim to assess correlations between MNBI and BI from HRIM (BI-HRIM) from 2 HRIM manufacturers in the setting of physiologic acid exposure time (AET).
Methods:
HRIM and pH-impedance monitoring demonstrating physiologic AET (< 4%) off proton pump inhibitors were required. BI-HRIM was extracted as the average from 5 cm and 10 cm above the lower esophageal sphincter. Distal BI-HRIM (DBI-HRIM) was also extracted from the most distal channel (Medtronic studies). MNBI was extracted from 6 channels. Concordance between BI-HRIM across manufacturers with MNBI was analyzed.
Results:
Thirty-six patients met the inclusion criteria (59.6 ± 1.7 years; 22% female; body mass index 30.5 ± 0.7; AET 1.6 ± 0.2%). Although MNBI was similar at all channels (P ≥ 0.18), Diversatek BI-HRIM was lower than Medtronic BI-HRIM (P = 0.003). Overall, BI-HRIM correlated with MNBI at corresponding recording sites, 7 cm and 9 cm (P < 0.05), but not at other sites (P≥ 0.19). Pearson’s correlations > 0.5 were seen at MNBI at 7 cm for both systems, and at 9 cm for Medtronic. DBI-HRIM correlated with MNBI at 3 cm and 5 cm (P < 0.03), but not at other locations (P > 0.1).
Conclusions
While numeric differences exist between manufacturers, BI-HRIM correlates with MNBI from corresponding channels in patients with physiologic AET. Comparison with AET elevation is needed to determine correlations between pathologic MNBI with BI-HRIM across manufacturers. The optimal HRIM channels from which BI values should be extracted also warrants further study.