1.Annual Report on the External Quality Assessment Scheme for Blood Bank Tests in Korea (2015).
Young Ae LIM ; Hyun Soo CHO ; Seog Woon KWON ; Kye Chul KWON ; Shin Young KIM ; Kyung Un PARK ; Hyun Jun PARK ; Woon Heung SONG ; Duck CHO ; Tae Hyoun UM ; Mee Kyung LEE ; Sang Moo HAN
Journal of Laboratory Medicine and Quality Assurance 2016;38(2):59-67
We report here the results of the external quality assessment scheme (EQA) of blood bank tests in Korea carried out in 2015. The proficiency testing specimens used in the survey were prepared at Ajou University Hospital. The response rates from participating laboratories for the first and second trials were 98.7% (542/549) and 98.2% (544/554), respectively. No answers to tests were considered incorrect, and the average accuracy rates for six different test items on the standard survey were as follows: ABO grouping, 99.4% to 100.0%; RhD typing, 99.4% to 100.0%; crossmatching, 93.6% to 99.0%; direct antiglobulin test (DAT) using a polyspecific reagent, 92.9% to 98.3%; DAT using an IgG monospecific reagent, 94.6% to 100.0%; DAT using a C3d monospecific reagent, 84.2% to 98.6%; unexpected antibody screening test, 94.5% to 100.0%; and antibody identification test, 93.8% to 100.0%. We performed a pilot survey on reactivities to A1 (54 responses) and H (50 responses); Rh C, c, E, and e antigen testing (47 responses); and ABO antibody titration (10-34 responses). We obtained excellent results for this EQA, and these results will be helpful for improving or maintaining the quality of the participating laboratories.
Blood Banks*
;
Coombs Test
;
Immunoglobulin G
;
Korea*
;
Laboratory Proficiency Testing
;
Mass Screening
2.Annual Report on External Quality Assessment of Blood Bank Tests in Korea (2013).
Young Ae LIM ; Jin Sook OH ; Seog Woon KWON ; Kye Chul KWON ; Shin Young KIM ; Kyung Un PARK ; Hyun Jun PARK ; Woon Heung SONG ; Duck CHO ; Tae Hyoun UM ; Kwang HUR ; Sang Moo HAN
Journal of Laboratory Medicine and Quality Assurance 2014;36(2):55-63
We report here the results of surveys for External Quality Assessment (EQA) of blood bank tests carried out in 2013. The proficiency testing specimens were prepared at Ajou University Hospital and sent to 548 and 545 institutes participating in the 1st and 2nd trial, respectively. Test items for the surveys were ABO grouping, RhD typing, crossmatching, direct antiglobulin test (DAT), antibody screening test, and antibody identification test. The response rates for the 1st and 2nd trials were 94.3% and 96.0%, respectively. No answers were considered incorrect answers, and the average accuracy rates of different test items of the survey were as follows: ABO grouping, 98.9% to 100%; RhD typing, 98.4% to 99.2%; crossmatching, 94.4% to 100.0%; DAT using polyspecific reagent, 94.5% to 99.7%; DAT using IgG monospecific reagent, 94.7% to 98.8%; DAT using C3d monospecific reagent, 91.3% to 98.6%; unexpected antibody screening test, 90.9% to 100%; and antibody identification test, 87.3% to 100.0%. Overall, we obtained excellent survey results for the EQA of blood bank tests carried out in 2013, and the number of participating institutes was higher in 2013 than in 2012.
Academies and Institutes
;
Blood Banks*
;
Coombs Test
;
Immunoglobulin G
;
Korea
;
Laboratory Proficiency Testing
;
Mass Screening
3.Annual Report on the External Quality Assessment Scheme for Blood Blank Tests in Korea (2014).
Young Ae LIM ; Jin Sook OH ; Seog Woon KWON ; Kye Chul KWON ; Shin Young KIM ; Kyung Un PARK ; Hyun Jun PARK ; Woon Heung SONG ; Duck CHO ; Tae Hyoun UM ; Mi Kyoung LEE ; Sang Moo HAN
Journal of Laboratory Medicine and Quality Assurance 2015;37(2):47-55
We report here the results of surveys on external quality assessment (EQA) of blood bank tests in Korea carried out in 2014. The proficiency testing specimens were prepared at Ajou University Hospital and the response rates for the 1st and 2nd trials were 94.3% (537/549) and 96.0% (545/554), respectively. No answers were considered incorrect, and the average accuracy rates of six different test items on the regular survey were as follows: ABO grouping, 98.5% to 100.0%; RhD typing, 98.1% to 99.4%; crossmatching, 91.2% to 99.6%; direct antiglobulin test (DAT) using a polyspecific reagent, 96.7% to 98.4%; DAT using an immunoglobulin-G monospecific reagent, 93.8% to 98.7%; DAT using a C3d monospecific reagent, 89.5% to 98.7%; unexpected antibody screening test, 96.2% to 100.0%; and antibody identification test, 69.8% to 100.0%. Test items for the pilot survey were reactivities to anti-A1 and anti-H, Rh subgrouping, and ABO antibody titration. Except for the result of the antibody identification test for specimens with multiple antibodies, we obtained excellent survey results for the EQA of blood bank tests carried out in 2014. In addition, the number of participating institutes was higher in 2014 than in 2013. The EQA of blood bank tests in 2014 should be helpful for improving the quality of the participating laboratories.