OBJECTIVE:To establish a method for the determination of related substances in Cabazitaxel injection. METHODS:HPLC method was used. The determination was performed on Agilent Eclipse XDB-C18column with mobile phase consisted of water-acetonitrile-ethanol(gradient elution)at the flow rate of 1.0 mL/min. The column temperature was 35 ℃,and the detection wavelength was set at 230 nm. The sample size was 20 μ L. Established method was used to determine related substances in 3 batches of Cabazitaxel injection. RESULTS:The linear relationship of cabazitaxel were 0.039-11.60 μ g/mL(r=0.999 8,n=7). The detection limit was 2×10-4μg,and quantitation limit was 8×10-4μg. RSD of precision and reproducibility tests were all lower than 10.0%(n=6). The amount of single impurity in 3 batches of samples ranged 0.07%-0.08%,and total amount of impurities were 0.26%-0.29%. CONCLUSIONS:Established method is simple,accurate and reliable,can be used for the determination of related substances in Cabazitaxel injection.